PubMed Health⌕ Search

PubMed · 15236462

A theoretical and experimental analysis of formulation and device parameters affecting solution MDI size distributions.

Abstract

The influence of formulation and device configurations on the initial droplet and residual particle size distribution from solution MDIs was theoretically and experimentally examined. Aerodynamic size distribution tests were conducted to characterize the size distribution of the residual particles formed when a solution MDI is actuated. The measured size distributions were approximately log-normally distributed, and did not show evidence of a secondary large particle mode. Theoretical relationships were developed to relate the residual particle size distribution to the initial size distribution of the atomized droplets. The residual particle size distribution was shown to be proportional to the nonvolatile concentration to the one-third power. Ethanol concentration, propellant type, valve size, and actuator orifice diameter were all shown to affect the initial droplet size distribution. Deposition of drug in the mouthpiece and USP inlet affect the measured size distribution during aerodynamic particle size measurements. Although there is a significant increase in the size of initial droplets as ethanol concentration increases, there is only a minor increase in the size of the residual particles measured when the USP Inlet is used due to size dependent deposition in the USP inlet and actuator mouthpiece. Vapor pressure was shown to explain only part of the differences in the size of the atomized droplets for various formulations. Theoretical and empirical equations were developed that make it possible to predict the residual particle size distribution for solution MDIs.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Stephen W Stein, Paul B Myrdal. 2004. A theoretical and experimental analysis of formulation and device parameters affecting solution MDI size distributions.. https://doi.org/10.1002/jps.20116

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Alternative statistical parameter for high-throughput screening assay quality assessment.

High-throughput screening is an essential process in drug discovery. The ability to identify true active compounds depends on the high quality of assays and proper analysis of data. The Z factor, presented by Zhang et al. in 1999, provides an easy and useful summary of assay quality and has been a widely accepted standard. However, as data analysis has undergone much improvement recently, the assessment of assay quality has not evolved in parallel. In this article, the authors study the implications of Z factor values under different conditions and link the Z factor with the power of discovering true active compounds. They discuss the different interpretations of Z factor depending on error distributions and advocate direct analysis of power as assay quality assessment. They also propose that in estimating assay quality parameters, adjustments in data analysis should be taken into account. Studying the power of identifying true "hits" gives a more direct interpretation of assay quality and may provide guidance in assay optimization on some occasions.

Chemistry, Pharmaceutical↗

Nonaqueous capillary electrophoresis in pharmaceutical analysis.

This review presents different solvents and electrolytes commonly used as BGEs in NACE for the analysis of pharmaceutical compounds. Most NACE applications carried out since 1998 for the analysis of compounds of pharmaceutical interest are presented in four tables: (i) analysis of drugs and related substances, (ii) analysis of chiral substances, (iii) analysis of phytochemical extracts and (iv) analysis of drugs in biological fluids. These selected examples are used to illustrate the interest in NACE versus conventional aqueous CE.

Chemistry, Pharmaceutical↗

Recent progress in capillary ITP.

ITP has been attracting constant attention for many years due to its principal capability to concentrate trace analytes by several orders of magnitude. In the current capillary format, it is able to concentrate trace analytes diluted to several microliters of an original sample into concentrated zones having volumes in the range of picoliters. Due to this reason, ITP holds an important position in many current multistage and multidimensional separation schemes. This article links up previous reviews on the topic and summarizes the progress of analytical capillary ITP since 2002. Almost 100 papers are reviewed that include methodological novelties, instrumental aspects, and analytical applications. Papers using ITP and/or isotachophoretic principles as part of multistage and/or multidimensional separation schemes are also included.

Chemistry, Pharmaceutical↗