Computer-assisted total knee replacement. A controlled cadaver study using a multi-parameter quantitative CT assessment of alignment (the Perth CT Protocol).

Abstract

A controlled study, comparing computer- and conventional jig-assisted total knee replacement in six cadavers is presented. In order to provide a quantitative assessment of the alignment of the replacements, a CT-based technique which measures seven parameters of alignment has been devised and used. In this a multi-slice CT machine scanned in 2.5 mm slices from the acetabular roof to the dome of the talus with the subject's legs held in a standard position. The mechanical and anatomical axes were identified, from three-dimensional landmarks, in both anteroposterior and lateral planes. The coronal and sagittal alignment of the prosthesis was then measured against the axes. The rotation of the femoral component was measured relative to the transepicondylar axis. The rotation of the tibial component was measured with reference to the posterior tibial condyles and the tibial tuberosity. Coupled femorotibial rotational alignment was assessed by superimposition of the femoral and tibial axial images. The radiation dose was 2.7 mSV. The computer-assisted total knee replacements showed better alignment in rotation and flexion of the femoral component, the posterior slope of the tibial component and in the matching of the femoral and tibial components in rotation. Differences were statistically significant and of a magnitude that support extension of computer assistance to the clinical situation.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

S K Chauhan, G W Clark, S Lloyd, R G Scott, W Breidahl, J M Sikorski. 2004. Computer-assisted total knee replacement. A controlled cadaver study using a multi-parameter quantitative CT assessment of alignment (the Perth CT Protocol).. https://doi.org/10.1302/0301-620x.86b6.15456

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

[Dislocation of artificial cartilage (SaluCartialge)].

Treament of focal cartilage defects is challenging, and techniques of bone marrow stimulation or cartilage transplantation are not sufficient to reconstitute hyaline cartilage. Preliminary results of artificial cartilage repair are promising, but no long-term results are available. Here, we report on a case of graft dislocation 6 months after artificial cartilage repair (SaluCartilage). We conclude that this method is not fully developed and more long-term results are required.

Arthroplasty, Replacement, Knee↗

[Osteotomy at knee level for young patients with gonarthrosis].

Young patients with gonarthrosis that does not respond adequately to conservative therapy can be treated by corrective osteotomy. Osteoarthritis of one compartment more often has a mechanical aetiology than osteoarthritis of the entire knee. Patients with osteoarthritis of the medial compartment often have a genu varum (bow-legs) while patients with osteoarthritis of the lateral compartment often have a genu valgum (knock-knees). The goal of corrective osteotomy is to transfer the load bearing to the normal compartment, which will reduce the symptoms and permit arthroplasty to be postponed. In retrospective studies, the procedure resulted in less pain, improved knee function or postponement of knee arthroplasty in 28-87% of the patients. Possible complications include pseudarthrosis, thromboembolism, contracture of the patellar tendon, paresis of the N. peroneus, compartment syndrome. The outcome of osteotomy for gonarthrosis depends on careful patient selection, the stage of osteoarthritis, and the achievement and maintenance of the correction of the load axis that was calculated before the operation.

Arthroplasty, Replacement, Knee↗