Current injection electrodes for electrical impedance tomography.
Abstract
Current conveyors have been identified as a possible component within the current injection electrodes of an electrical impedance tomography system, where accurate current generation or precise measurement of the current injected is required. Several circuit configurations have been investigated through simulation to determine the most suitable to meet the specifications of the EIT system. A bipolar (floating source) circuit configuration employing the use of current conveyors has been designed, which achieves greater than 12 mA output current without saturation, over an accepted body impedance range. Simulations were performed over frequencies in excess of 1 MHz, and the output phase shift was less than 0.15 degrees up to 250 kHz, and 0.6 degrees up to 1 MHz.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
S Armstrong, D Jennings. 2004. Current injection electrodes for electrical impedance tomography.. https://doi.org/10.1088/0967-3334%2F25%2F4%2F001