PubMed · 15541680

Choosing a cervical disc replacement.

Abstract

BACKGROUND CONTENT: Three important basic scientific studies are presented that measured the volumetric density of longitudinal bony columns within the cervical vertebra. The most solid bone is lateral, adjacent to the uncovertebral joints in a radial pattern. PURPOSE: To characterize the best footprint, profile and biomaterials to construct a cervical disc replacement. STUDY DESIGN: A compilation of biomechanical and anatomical basic scientific studies. METHODS: Microcomputed tomographic imaging, trabecular density and mineral distribution were quantitated from human cervical vertebra. RESULTS: The lateral portions of the cervical vertebra are subjected to higher bending loads than the lumbar vertebral bodies. Therefore, the optimal prosthesis needs to be anchored in the lateral uncovertebral bone. To reduce the incidence of cervical subsidence, the prosthesis needs to be more rectangular than round to take advantage of the radially oriented lateral trabeculae. TiCaP (titanium/calcium phosphate) (Cervitech, Inc., Rockaway, NJ) bony ingrowth coating leads to 10% to 15% greater bony integration than plasma-sprayed titanium. TiCaP causes a supersaturated solution of CaP at the metal-bone interface, which enables reprecipitation of hydroxyapatite and superior bony integration. The optimum pore size of the ingrowth coating of the lumbar spine is 75 to 300 microns, whereas in the cervical spine the optimal ingrowth coating is 20 to 30 microns. This is an order of magnitude lower in pore size to match the smaller cervical trabecular architecture. CONCLUSIONS: Kinematic considerations for the cervical spine show the load is 1/9th the load carried by the lumbar spine or 50 N per segment. Knowing the sliding distance and wear characteristics of conventional biomaterials (ultrahigh molecular weight polyethylene and cobalt chrome) demonstrates that the generation of particulate debris should be a very minor consideration with cervical arthroplasty.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Helmut D Link, Paul C McAfee, Luiz Pimenta. Choosing a cervical disc replacement.. https://doi.org/10.1016/j.spinee.2004.07.022

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Guidelines for the replacement of temporomandibular joints in the United Kingdom.

BAOMS has requested that guidelines be formulated for the replacement of the temporomandibular joint (TMJ). This is an expensive and technique sensitive method of TMJ reconstruction and in the current climate warrants an agreed approach. The following document states the indications and contraindications for this technique as discussed and agreed amongst surgeons currently carrying out this procedure in the UK.

Arthroplasty, Replacement↗

Influence of changes in the composition of an acrylic bone cement on its polymerization kinetics.

It has been suggested in the literature that a lower polymerization rate of an acrylic bone cement is favorable for the in vivo longevity of a cemented arthroplasty. The present work was a study of the influence of three changes in the composition of an acrylic bone cement (when taken separately) on the cement polymerization rate at 37 degrees C (assumed to be the temperature in the bone bed during a cemented arthroplasty) [k']. The changes were the amount of copolymer as a proportion of the total powder weight (in cements in which there is a copolymer in the powder), the amount of DMPT as a proportion of the total volume of the liquid monomer, and the accelerator. k' was calculated using values of the activation energy and the frequency factor (assuming the polymerization reaction is Arrhenius in nature) that were computed from measurements made using the nonisothermal mode of differential scanning calorimetry. Statistical analysis (one-way ANOVA, with Bonferroni correction, and factorial ANOVA) of the k' values showed that the change in accelerator had a significant influence on k'. The importance of this finding, together with results from two relevant literature reports, is discussed within the context of the use of modified bone cements in cemented arthroplasties.

Arthroplasty, Replacement↗

Glenohumeral kinematics following total shoulder arthroplasty: a finite element investigation.

The osseous geometry of the glenohumeral joint is naturally nonconforming and minimally constrained, and the joint's stability is maintained by action of the rotator cuff muscles. Damage to these muscles is often associated with joint degeneration, and a variety of glenoid prostheses have been developed to impart varying degrees of stability postoperatively. The issues of conformity and constraint within the artificial shoulder have been addressed through in vivo and in vitro studies, although few computational models have been presented. The current investigation presents the results of three-dimensional finite element analyses of the total shoulder joint and the effects of design parameters upon glenohumeral interaction. Conformity was shown not to influence the loads required to destabilize the joint, although it was the principal factor determining the magnitude of humeral head translation. Constraint was found to correlate linearly with the forces required to dislocate the humeral head, with higher constraint leading to slightly greater humeral migration at the point of joint instability. The model predicts that patients with a dysfunctional supraspinatus would experience frequent eccentric loading of the glenoid, especially in the superior direction, which would likely lead to increased fixation stresses, and hence, a greater chance of loosening. For candidates with an intact rotator cuff, the models developed in this study predict that angular constraints of at least 14 degrees and 6.5 degrees in the superoinferior and anteroposterior axes are required to provide stable unloaded abduction of the humerus, with larger constraints of 18 degrees and 10 degrees necessitated by a dysfunctional supraspinatus. The tools developed during this study can be used to determine the capacity for different implant designs to provide resistance to excessive glenohumeral translations and reduce the potential for instability of the joint, allowing surgeons to optimize postoperative functional gains on a patient by patient basis.

Arthroplasty, Replacement↗