Equipotential earthing study: equipotential wiring is unnecessary.
Abstract
This study was undertaken to determine the ongoing need for Equipotential (EP) earthing as required in AS/NZS 3003, section 4.4.2. A total of 2279 earth points were tested in four clinical institutions in Body Protected areas (as defined in AS/NZS 3003). Each earth point was tested for potentials existing in the earthing in order to assess whether any potentials exceeded the 100 mV requirement of AS/NZS 3003 for EP earthing. The voltage measured was 1.21 mV +/- 3.59 mV, (mean +/- standard deviation (SD)), with a of 99 % confidence interval between 1.0 mV and 1.4 mV. This study suggests that the requirements for EP earthing in section 4.4.2 of AS/NZS 3003 are unnecessary. It is further suggested that Standards Australia review these requirements with the view of withdrawing them from the Standard.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
K Pui-tse Foong, A Watson, A Avolio, Standards Australia. 2004. Equipotential earthing study: equipotential wiring is unnecessary.. https://doi.org/10.1007/bf03178675