Sphygmomanometers--an audit in general practice.

Abstract

BACKGROUND: The accuracy of sphygmomanometers used in Australian general practice is unknown but potentially important. METHOD: We measured the accuracy of sphygmomanometers in general practice in the Hunter region of New South Wales using a gold standard. Practices were recruited by an advertisement in the division newsletter. RESULTS: Sixty practices (35%) volunteered. A total of 404 instruments were checked. Over 95% of sphygmomanometers were within 4 mmHg of gold standard sphygmomanometer across the clinical pressure range. Mercury sphygmomanometers were more accurate than aneroid (p<0.01). There was no significant association between accuracy and age, calibration, or visual inspection of the instruments. DISCUSSION: This study demonstrated a high accuracy rate of the sphygmomanometers checked, especially those sphygmomanometers that were mercury.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Nayankumar C Shah, David W Sibbritt, Susan Heaney, Jan Sharples. 2004. Sphygmomanometers--an audit in general practice.. https://pubmed.ncbi.nlm.nih.gov/15584340/

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Tip cooling effect and failure mechanism of field-emitting carbon nanotubes.

The cooling effect accompanying field electron emission has been considered for a single carbon nanotube (CNT) used as a field emission (FE) electron source. An improved model for the failure mechanism of field emitting CNTs has been proposed and validated. Our model predicts a maximum temperature (T-max) located at an interior point rather than the tip of the CNTs, and the failure of the CNT emitters tends to take place at the T-max point, inducing a segment by segment breakdown process. A combination of Joule heating and electrostatic force effect is proposed responsible for initiating the failure of the field emitting CNT and validated by in situ FE observation.

Equipment Failure↗

[Importance of an initial check-up for diagnostic radiology apparatus management].

The management of apparatus for diagnostic imaging is an important job for radiological technologists. In the last 10 years we have encountered 2227 malfunctions in 3652 days. We detected 163 cases at the initial check-up, which accounted for 7.3% of total cases. By performing an initial check-up, we detected one malfunction every 21 days and were able to keep the effects of mishaps to a minimum, prevent accidents, and obtain stable-quality images.

Equipment Failure↗