PubMed · 1569868
Perfusion and diffusion MR imaging.
Abstract
Diffusion-weighted images of the rat brain were obtained using the pulsed-gradient spin-echo method. An attempt was made to extract perfusion-related parameters from signal intensity data taken from the caudate-putamen region of the images, by using a nonlinear least-squares calculation to fit the Le Bihan biexponential expression (Le Bihan et al., Radiology, 168, 497 (1988)) to the data. The perfusion-related parameters could not be obtained with sufficient accuracy to be useful, although the perfusion-weighted images appear to contain meaningful qualitative information. An analysis of the perfusion model is presented and shows why the Le Bihan pseudo-diffusion coefficient is particularly difficult to measure with reasonable accuracy.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
M D King, N van Bruggen, A L Busza, J Houseman, S R Williams, D G Gadian. 1992. Perfusion and diffusion MR imaging.. https://doi.org/10.1002/mrm.1910240210