High-frequency oscillatory ventilation: lessons learned from mechanical test lung models.

Abstract

OBJECTIVE: Review data obtained from high-frequency oscillatory ventilation (HFOV) and mechanical test lung models with respect to delivered tidal volume, distal pressure transmission, endotracheal tube cuff leaks, and simulated clinical conditions. DESIGN: Review of selected studies from PubMed, published abstracts, and institutional mechanical test lung data. RESULTS: Tidal volume delivery during HFOV is altered by oscillatory pressure amplitude (DeltaP), frequency (Hz), percent inspiratory time (IT%), and patient variables. Distal (carinal) oscillatory pressure amplitude transmission is directly correlated with endotracheal tube diameter and peripheral airway resistance. Endotracheal tube cuff leaks promote egress of tracheal gas while attenuating distal oscillatory pressure amplitude and tidal volume transmission. Simulated clinical conditions (e.g., increased distal airway resistance, mainstem intubation) may increase observed DeltaP, whereas mean airway pressure is decreased with air leaks. CONCLUSION: Mechanical test lung and artificial trachea simulations may provide useful information on the interaction of HFOV with altered lung mechanics and may contribute to the formulation of HFOV clinical strategies. Important limitations of these models include absence of gas exchange, histologic and biologic markers, or hemodynamic data.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Michael Van de Kieft, David Dorsey, David Morison, Lazaro Bravo, Steven Venticinque, Stephen Derdak. 2005. High-frequency oscillatory ventilation: lessons learned from mechanical test lung models.. https://doi.org/10.1097/01.ccm.0000155924.74942.7f

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Sequential meta-analysis of past clinical trials to determine the use of a new trial.

BACKGROUND: Clinical trials can be stopped early based on interim analyses or sequential analyses. In principle, sequential analyses can also be used to decide whether enough evidence has been gathered in completed trials to make further trials unnecessary. We demonstrate such an application through a retrospective analysis of clinical trials comparing ventilation methods for the treatment of preterm newborns. METHODS: We identified 5 recent trials that compared high-frequency ventilation with conventional mechanical ventilation in the treatment of preterm newborns. Death or chronic lung disease and chronic lung disease in survivors were the primary clinical outcomes of interest. We applied sequential meta-analyses to these 5 studies. RESULTS: After including the first study of the last 5 trials in a sequential meta-analysis, the boundary of "no clinically relevant effect" was crossed for both outcomes (death or chronic lung disease). A sensitivity analysis using a reduction in the size of assumed clinically relevant effect showed the same findings after 2 trials. CONCLUSIONS: Sequential meta-analyses showed that a lack of clinically relevant effect had been established after the first of the 5 trials. If such an analysis had been conducted after the first or second of these clinical trials, it might have led to changes in the study design of subsequent trials or even to a reassessment of the need for further trials.

High-Frequency Ventilation↗