PubMed · 15798340
Modelling simple helically delivered dose distributions.
Abstract
In a previous paper, we described quality assurance procedures for Hi-Art helical tomotherapy machines. Here, we develop further some ideas discussed briefly in that paper. Simple helically generated dose distributions are modelled, and relationships between these dose distributions and underlying characteristics of Hi-Art treatment systems are elucidated. In particular, we describe the dependence of dose levels along the central axis of a cylinder aligned coaxially with a Hi-Art machine on fan beam width, couch velocity and helical delivery lengths. The impact on these dose levels of angular variations in gantry speed or output per linear accelerator pulse is also explored.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
John D Fenwick, Wolfgang A Tomé, Michael W Kissick, T Rock Mackie. 2005-03-22. Modelling simple helically delivered dose distributions.. https://doi.org/10.1088/0031-9155%2F50%2F7%2F013
Cite the original work for its findings. Save a collection to share your selection of sources.