PubMed · 1587106
Electrical impedance tomography systems based on voltage drive.
Abstract
At the ESAT-MICAS research facility in Leuven, several EIT systems have been designed and realised. The latest hardware set-up makes use of a PC to control the data collection and to reconstruct the images. A voltage drive strategy and no common-mode feedback are some of its specific characteristics. The function-generator produces signals with a frequency between 10 and 100 kHz, so multifrequency images can also be produced. Static images have already been obtained and (semi-)real time imaging is possible with our latest mark IV system. This system has 16 bit analogue-to-digital convertors and is capable of taking 50 x 10(3) samples/s.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
W Sansen, B Geeraerts, W Van Petegem, M Steyaert. 1992. Electrical impedance tomography systems based on voltage drive.. https://doi.org/10.1088/0143-0815%2F13%2Fa%2F008
Cite the original work for its findings. Save a collection to share your selection of sources.