PubMed Health⌕ Search

PubMed · 15889434

Analysis of sterilization protocols for peptide-modified hydrogels.

Abstract

Concerns about the efficacy of ethanol disinfection for implanted biomaterials prompted investigation of an alternative sterilization process, ultraviolet irradiation, for terminal sterilization of N-isopropylacrylamide-based hydrogels containing biomimetic peptides. Ultraviolet irradiation is more easily applied on a laboratory scale than gamma irradiation or electron beam, two commercially utilized methods; thus, UVC irradiation was investigated as a low-cost sterilization procedure that might be performed in laboratories prior to in vivo studies. UVC irradiation at 400 muW/cm(2) for up to 15 h did not prevent growth of Escherichia coli within the hydrogels, while ethanol disinfection did prevent growth for the duration of the experiment (120 h). Furthermore, UVC irradiation caused progressive degradation of peptides containing the Arg-Gly-Arg (RGD) domain. UVC irradiation cannot be used as a terminal sterilization process for peptide-modified materials. The system used in this study is not intended to be adequate for evaluating the sterility of medical devices in accordance with current Good Manufacturing Practice (cGMP); however, it remains a useful, low-cost system for the preliminary evaluation of sterilization procedures in terms of their ability to eliminate pathogenic organisms while preserving the structure of biologically active molecules within in a laboratory setting. Ethanol treatment is still the preferred method for disinfection of bioactive materials containing peptides or UV-degradable groups.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Nathaniel Huebsch, Michele Gilbert, Kevin E Healy. 2005. Analysis of sterilization protocols for peptide-modified hydrogels.. https://doi.org/10.1002/jbm.b.30155

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Ion-selective channels in K562 cells: a patch-clamp analysis.

Four types of ion-selective channels were found by the patch-clamp technique in the human erythroleukemia K562 cells. I) in cell-attached configuration at potentials less negative than -40 mV an 8 ps channel was detected. The potential dependence of channel activity suggests that this is the TTX-sensitive Na+ channel. II) A cation-selective channel was observed with equal permeability for Na+ and K+ and a potential-independent single-channel conductance of 19 pS. The channel is activated by intracellular Ca2+ and inhibited by TEA, III) A predominantly anion-selective channel was identified with the selectivity sequence NO3- > J- > Cl- = Br- >> SO4(2-). The single-channel conductance shows outward rectification, and is in symmetrical NaCl solution 19 pS at -60 mV and 54pS at +50 mV. The open- and closed-time distributions suggest one open and at least four closed states. At submicromolar concentrations, the open state is blocked by H2DIDS leading to channel flicker between open and blocked channel; higher concentrations (apparent KI = 6.8 uM) lead to a longer-lasting blocked state. Both components of inhibition are reversible. IV) In addition, an 8 pS, Na(+)- and K(+)- selective channel could be induced by application of palytoxin. For channel activity, the presence of extracellular Na+ is essential. It is assumed that the Na+, K(+)-pump molecule is involved in the channel formation. Similarly, it is discussed whether the anion-selective channel represents a pore conformation of an electrically silent anion exchanger.

Acrylamides↗

In-situ AFM studies of the phase-transition behavior of single thermoresponsive hydrogel particles.

The volume phase transition (VPT) behavior of individual thermally responsive poly(N-isopropylacrylamide-co-acrylic acid) (pNIPAm-co-AAc) hydrogel microparticles was studied by in-situ dynamic mode atomic force microscopy (AFM) and force spectroscopy during heating and cooling cycles. Hydrogel samples were prepared by electrostatic immobilization of microparticles to amine-modified gold surfaces. The AFM studies of particle deswelling were performed by varying the force applied on the particles during imaging as a function of the geometry and material of the AFM probe. Aluminum-coated silicon cantilevers were found to influence substantially the behavior of the particles during the VPT, leading to a significant shape change. Low force impact magnetic excitation of the AFM probe (MAC mode) during dynamic mode measurements resulted in an undisturbed deswelling behavior enabling observation of the expected volume changes of the particles without significant tip-sample interaction. Hence, MAC-mode AFM was determined to be the most suitable technique for in-situ AFM studies on volume and shape changes at single hydrogel particles during VPT. Elasticity measurements performed at single particles at temperatures below and above the VPT revealed a 15-fold increase in the Young's modulus after passing the VPT, indicating the transition from a soft, swollen network to a stiffer, deswollen state.

Acrylamides↗

Synthesis and characterization of variable-architecture thermosensitive polymers for complexation with DNA.

Copolymers of N-isopropylacrylamide with a fluorescent probe monomer were grafted to branched poly(ethyleneimine) to generate polycations that exhibited lower critical solution temperature (LCST) behavior. The structures of these polymers were confirmed by spectroscopy, and their phase transitions before and after complexation with DNA were followed using ultraviolet and fluorescence spectroscopy and light scattering. Interactions with DNA were investigated by ethidium bromide displacement assays, while temperature-induced changes in structure of both polymers and polymer-DNA complexes were evaluated by fluorescence spectroscopy, dynamic light scattering, laser Doppler anemometry, and atomic force microscopy (AFM) in water and buffer solutions. The results showed that changes in polymer architecture were mirrored by variations in the architectures of the complexes and that the overall effect of the temperature-mediated changes was dependent on the graft polymer architecture and content, as well as the solvent medium, concentrations, and stoichiometries of the complexes. Furthermore, AFM indicated subtle changes in polymer-DNA complexes at the microstructural level that could not be detected by light scattering techniques. Uniquely, variable-temperature aqueous-phase AFM was able to show that changes in the structures of these complexes were not uniform across a population of polymer-DNA condensates, with isolated complexes compacting above LCST even though the sample as a whole showed a tendency for aggregation of complexes above LCST over time. These results indicate that sample heterogeneities can be accentuated in responsive polymer--DNA complexes through LCST-mediated changes--a factor that is likely to be important in cellular uptake and nucleic acid transport.

Acrylamides↗