PubMed · 1591865
Serum analysis for potassium ions using a fibre optic sensor.
Abstract
A fibre optic potassium ion sensor has been developed based on reflectance measurements of the pH-dependent chromogenic crown ether 'Takagi reagent' immobilized on the non-ionic polymeric resin Amberlite XAD2. The sensor response is in the 1.5-100 mM range with a detection limit of 0.25 mM K+. The selectivity factor over sodium is 20. The proposed sensor works with a reproducibility of +/- 0.8% at the 5 mM K+ level. Advantageous design features include ease of construction, simplicity of use, reversibility, short response time (approximately 1 min) and an optimum working pH range (7.0-8.0) for biological fluids. This optical sensor was applied successfully to the direct determination of potassium in human serum. Aqueous potassium standards containing 0.13 M NaCl and a buffer of pH 8 were used for calibration.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
R Pereiro García, F Alava Moreno, M E Díaz García, A Sanz-Medel, R Narayanaswamy. 1992-04-30. Serum analysis for potassium ions using a fibre optic sensor.. https://doi.org/10.1016/0009-8981(92)90148-j