Volume-based characterization of postocclusion surge.

Abstract

PURPOSE: To propose an alternative method to characterize postocclusion surge using a collapsible artificial anterior chamber to replace the currently used rigid anterior chamber model. SETTING: Fundación Oftamológica Los Andes, Santiago, Chile. METHODS: The distal end of a phacoemulsification handpiece was placed inside a compliant artificial anterior chamber. Digital recordings of chamber pressure, chamber volume, inflow, and outflow were performed during occlusion break of the phacoemulsification tip. The occlusion break profile of 2 different consoles was compared. RESULTS: Occlusion break while using a rigid anterior chamber model produced a simultaneous increase of chamber inflow and outflow. In the rigid chamber model, pressure decreased sharply, reaching negative pressure values. Alternatively, with the collapsible chamber model, a delay was observed in the inflow that occurs to compensate the outflow surge. Also, the chamber pressure drop was smaller in magnitude, never undershooting below atmospheric pressure into negative values. Using 500 mm Hg as vacuum limit, the Infiniti System (Alcon) performed better that the Legacy (Alcon), showing an 18% reduction in peak volume variation. CONCLUSIONS: The collapsible anterior chamber model provides a more realistic representation of the postocclusion surge events that occur in the real eye during cataract surgery. Peak volume fluctuation (mL), half volume recovery time(s), and volume fluctuation integral value (mL x s) are proposed as realistic indicators to characterize the postocclusion surge performance. These indicators show that the Infiniti System has a better postocclusion surge behavior than the Legacy System.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Jaime Zacharias, Sergio Zacharias. 2005. Volume-based characterization of postocclusion surge.. https://doi.org/10.1016/j.jcrs.2005.03.061

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Model of pulsatile-flow of aqueous humor through the iris-lens canal.

PURPOSE: To present a model of pulsatile-flow of aqueous humor from posterior (PC) to anterior chamber (AC) and to analyze the sensitivity of this novel model in detecting typical high risk conditions predisposing to pupillary block. METHODS: The model assumes noncontinuous flow of aqueous through the iris-lens canal. Aqueous that fills the canal will be ejected toward the AC-side of the canal at certain time intervals, and between 2 events of aqueous ejection there is no actual flow through this canal. Pupillary pumping rate (PPR) was calculated from the aqueous flow rate and the calculated volume of iris-lens canal. RESULTS: PPR values were generated by incorporating pupillary diameter (1 to 8 mm), aqueous flow rate (1 to 2.5 microL/min), and iris-lens canal width (0.5 to 2 mm) and height (3-9 microm) in numerical experimentation with the present model. PPR showed inverse dependence on iris-lens canal height and pupillary diameter and was directly proportional to aqueous flow rate, in agreement with the steady-flow model. However, contrary to the steady-flow model, PPR showed inverse dependence on iris-lens canal width and predicted the anticipated PC-AC pressure gradient changes at simulated light-dark transition in eyes of patients with clinically narrow angles and ultrasound biomicroscopy evidenced pupillary block. CONCLUSIONS: Upon the incorporation of real ultrasound biomicroscopy measurements in numerical experimentations with both models, the present pulsatile-flow model, contrary to the steady-flow model, showed good predictability of PC-AC pressure gradient changes in a typical condition predisposing to pupillary block.

Anterior Chamber↗