PubMed Health⌕ Search

PubMed · 16398721

Thresholds for significant decrease in hemodialysis access blood flow.

Abstract

During hemodialysis access surveillance, referral for evaluation and correction of stenosis is based upon determination that a significant decrease in blood flow (Q) has occurred. However, criteria for determining when a decrease is statistically significant have not yet been established. In this study we established such criteria by analyzing Q variation with the glucose pump test (GPT). We took nine Q measurements in each of 25 patients (18 grafts, 7 fistulas) during three dialysis sessions within a 2-week period (predialysis and during hours 1 and 3). We determined thresholds that define a significant percentage decrease in Q (deltaQ) for various p values. In order to confirm the general applicability of these thresholds, we computed the average within-patient Q variation during the three sessions (computed as a coefficient of variation and referred to as short-term variation). We then determined the relative influences of biological (true) variation and analytical error on short-term variation. We found that deltaQ must be > 33% to be significant at p < 0.05, whereas the threshold is > 17% for p < 0.20. Measuring Q at uniform versus different times during the sessions did not significantly reduce these thresholds. We also found that biological variation was nearly as large as short-term Q variation, whereas analytical error contributed minimally to short-term variation. In conclusion, this study defines thresholds for a significant deltaQ that have wide application in determining access referral for evaluation and correction of stenosis. Selection of a particular threshold should consider the relative importance of avoiding thrombosis versus avoiding unnecessary procedures. If avoiding unnecessary procedures is a priority, then we recommend a threshold of > 33%. These thresholds apply to other methods of measuring Q, provided analytical error is significantly less than biological variation.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Sunanda J Ram, Raja Nassar, Rashid Sharaf, Alberto Magnasco, Steven A Jones, William D Paulson. Thresholds for significant decrease in hemodialysis access blood flow.. https://doi.org/10.1111/j.1525-139x.2005.00104.x

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Pre-operative radial arterial diameter predicts early failure of arteriovenous fistula (AVF) for haemodialysis.

INTRODUCTION: Long term patency of arteriovenous fistula (AVF) is relevant to the management of end stage renal failure (ESRF) patients on haemodialysis (HD). We evaluated the role of routine radial arterial duplex for imaging radial artery before AVF formation to investigate the relationaship between radial artery internal diameter (ID) and AVF patency. METHODS: 21 patients with ESRF were examined by duplex sonography before AVF formation, 1 day, 1 week, 4 week and 12 weeks post AVF formation. For assessment of AVF patency, patients were divided into 2 groups. Group-1, 11 patients with radial artery ID <1.5mm and Group-2, 10 patients with radial artery ID >1.5mm. Measurement of radial artery blood inflow rate was calculated from mean blood flow velocity and vessel diameter. All AVFs were constructed on the forearm using autologous veins. RESULTS: In Group-1, 5 patients (45%) showed immediate thrombosis of AVF graft. All patients in group-2 had patent AVF at 12 weeks. Pre-AVF formation radial artery blood inflow rate between two groups was not significantly different (p=0.06). Radial artery blood inflow rate was consistently and significantly higher in group-2 at all later time points with p value of <0.01 (Mann Whitney test). CONCLUSION: There was a high failure rate of AVF with radial artery ID of <1.5mm. In the presence of small radial arteries primary access AVF in the upper arm should be considered.

Arteriovenous Shunt, Surgical↗

Corlett loop for microvascular reconstruction in a neck depleted of vessels.

Free tissue transfer has revolutionised reconstruction of the head and neck and there are currently few local factors that preclude its use. These include a lack of suitable recipient vessels in the area for microvascular anastomosis. We describe the use of a Corlett loop in which a temporary arteriovenous fistula (AVF) is created with a transposed cephalic vein to enable reconstruction of a mandibular defect with a fibular free flap.