PubMed · 16642991
Biexponential fitting of diffusion-ordered NMR data: practicalities and limitations.
Abstract
High resolution diffusion-ordered NMR spectroscopy (HR-DOSY) generally uses monoexponential fitting of the diffusional attenuation of pulsed-field gradient stimulated echo spectra and can, thus, only give proper separation of signals in the diffusion domain where the individual peaks are well-resolved in the spectral domain. In principle, it should be possible to resolve the decays of coincident spectral peaks by multiexponential fitting, but the well-known fundamental difficulties in such a separation are exacerbated by instrumental distortions of the signal decay. The limitations imposed on biexponential fitting by finite signal-to-noise ratio and by the distortion of the theoretical form of the signal attenuation as a result of pulsed-field gradient nonuniformity are explored, and the improvement in DOSY spectra produced using biexponential fitting afforded by compensating for the latter are illustrated.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Mathias Nilsson, Mark A Connell, Adrian L Davis, Gareth A Morris. 2006-05-01. Biexponential fitting of diffusion-ordered NMR data: practicalities and limitations.. https://doi.org/10.1021/ac060034a