PubMed · 16644977
Ferrous-Ferric Ion exchange dosemeter.
Abstract
In this work a three-dimensional ferrous-ferric ion exchange dosemeter is proposed and the dose response measured. The dosemeter consists of strong acid cation exchange resin beads in the H form in water. Amberlyst 15 Wet beads with a harmonic mean diameter of 0.600-0.850 mm were prepared by soaking them in an aqueous solution of ferrous ammonium sulphate to exchange ferrous ions for H(+) ions. The beads were rinsed with distilled water and packed in glass vials. Sets of samples with ferrous ion concentrations of 0.5 and 1.0 mM were dosed with 6 MV X rays from a Varian 2100C linac. The spin-lattice relaxation time constants (T1) for the samples were measured using an Apollo spectrometer (Tecmag, Houston, TX) interfaced to a 1.5 T magnet (Magnex, Abingdon, UK). Each sample had two T1 values; a long T1 at 1200 ms that did not significantly change with dose and a short T1 that ranged from 56 ms at 0 Gy to 36 ms at 100 Gy. The R1 vs. dose responses were linear with slopes of 0.066 and 0.079 s(-1) Gy(-1).
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
John A Bauhs, Bruce E Hammer. 2006-04-27. Ferrous-Ferric Ion exchange dosemeter.. https://doi.org/10.1093/rpd%2Fnci594
Cite the original work for its findings. Save a collection to share your selection of sources.