Alterations in surface geometry in retrieved polyethylene glenoid component.

Abstract

This study characterized and quantified the changes found in retrieved glenoid polyethylene components found at revision total shoulder arthroplasty (TSA). Twenty components obtained at revision TSA were evaluated, all from a system (Global, DePuy, Warsaw, IN) with a glenoid radius of curvature 3 mm greater than that of the humeral head. Laser surface scanning provided three-dimensional analysis of the surface of the glenoid component. Scans of unused components of similar sizes enabled determination of the changes occurring after implantation. Alterations in radius of curvature were noted in every glenoid. All showed loss of the balance stability angle (BSA, the maximal angle that the net humeral joint reaction force can make with the glenoid center line before the humeral head would dislocate) of at least 5 degrees in one or more directions. Increase in BSA in one direction was seen in 11 of the components. In five of these, the increase was associated with a reduction of the local radius to match that of the humeral component. Glenoid surface morphology and stability can be changed by in vivo use. While correlation with clinical instability in the patients from whom the implants were obtained was not possible, many of the observed changes in surface morphology are of sufficient magnitude to compromise the contribution of the glenoid surface to shoulder stability. Three patterns of wear were identified: "humeral" that showed loss of the mismatch between the humeral and glenoid radii of curvature (5 of 20 components), "diffuse" that showed broad surface irregularity (18 of 20), and "rim" wear with loss of the polyethylene rim of the component (14 of 20). More than one type of wear was possible within a single glenoid.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Jonathan P Braman, Alexis Falicov, Richard Boorman, Frederick A Matsen. 2006. Alterations in surface geometry in retrieved polyethylene glenoid component.. https://doi.org/10.1002/jor.20158

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Guidelines for the replacement of temporomandibular joints in the United Kingdom.

BAOMS has requested that guidelines be formulated for the replacement of the temporomandibular joint (TMJ). This is an expensive and technique sensitive method of TMJ reconstruction and in the current climate warrants an agreed approach. The following document states the indications and contraindications for this technique as discussed and agreed amongst surgeons currently carrying out this procedure in the UK.

Arthroplasty, Replacement↗

Influence of changes in the composition of an acrylic bone cement on its polymerization kinetics.

It has been suggested in the literature that a lower polymerization rate of an acrylic bone cement is favorable for the in vivo longevity of a cemented arthroplasty. The present work was a study of the influence of three changes in the composition of an acrylic bone cement (when taken separately) on the cement polymerization rate at 37 degrees C (assumed to be the temperature in the bone bed during a cemented arthroplasty) [k']. The changes were the amount of copolymer as a proportion of the total powder weight (in cements in which there is a copolymer in the powder), the amount of DMPT as a proportion of the total volume of the liquid monomer, and the accelerator. k' was calculated using values of the activation energy and the frequency factor (assuming the polymerization reaction is Arrhenius in nature) that were computed from measurements made using the nonisothermal mode of differential scanning calorimetry. Statistical analysis (one-way ANOVA, with Bonferroni correction, and factorial ANOVA) of the k' values showed that the change in accelerator had a significant influence on k'. The importance of this finding, together with results from two relevant literature reports, is discussed within the context of the use of modified bone cements in cemented arthroplasties.

Arthroplasty, Replacement↗

Glenohumeral kinematics following total shoulder arthroplasty: a finite element investigation.

The osseous geometry of the glenohumeral joint is naturally nonconforming and minimally constrained, and the joint's stability is maintained by action of the rotator cuff muscles. Damage to these muscles is often associated with joint degeneration, and a variety of glenoid prostheses have been developed to impart varying degrees of stability postoperatively. The issues of conformity and constraint within the artificial shoulder have been addressed through in vivo and in vitro studies, although few computational models have been presented. The current investigation presents the results of three-dimensional finite element analyses of the total shoulder joint and the effects of design parameters upon glenohumeral interaction. Conformity was shown not to influence the loads required to destabilize the joint, although it was the principal factor determining the magnitude of humeral head translation. Constraint was found to correlate linearly with the forces required to dislocate the humeral head, with higher constraint leading to slightly greater humeral migration at the point of joint instability. The model predicts that patients with a dysfunctional supraspinatus would experience frequent eccentric loading of the glenoid, especially in the superior direction, which would likely lead to increased fixation stresses, and hence, a greater chance of loosening. For candidates with an intact rotator cuff, the models developed in this study predict that angular constraints of at least 14 degrees and 6.5 degrees in the superoinferior and anteroposterior axes are required to provide stable unloaded abduction of the humerus, with larger constraints of 18 degrees and 10 degrees necessitated by a dysfunctional supraspinatus. The tools developed during this study can be used to determine the capacity for different implant designs to provide resistance to excessive glenohumeral translations and reduce the potential for instability of the joint, allowing surgeons to optimize postoperative functional gains on a patient by patient basis.

Arthroplasty, Replacement↗