An inductive power link for a wireless endoscope.

Abstract

An inductive link is presented that can handle freedom of motion. The envisaged application is to power a wireless camera capsule used for non-invasive visual inspection of the small bowel. Up to 150 mW of usable dc power can be delivered to the capsule for the entire duration of its travel along the gastric track. The outer dimensions of the power receiver compartment inside the capsule are Ø 10 mmx13 mm. The power efficiency of the link is measured to be 1% under worst-case geometrical conditions. The issue of patient's health and safety regarding exposure to the electromagnetic field is addressed. To this purpose, an experimental method is applied to predict specific absorption rates. Conversely, the effect of biological tissue on link performance and practical workability is evaluated as well. Electric shielding of the transmitter coil is applied to reduce biological tissue interaction. In this way, appliance on a living subject is made both safe and practically feasible.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Bert Lenaerts, Robert Puers. 2006-08-14. An inductive power link for a wireless endoscope.. https://doi.org/10.1016/j.bios.2006.06.015

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Wireless capsule endoscopy: from diagnostic devices to multipurpose robotic systems.

In the recent past, the introduction of miniaturised image sensors with low power consumption, based on complementary metal oxide semiconductor (CMOS) technology, has allowed the realisation of an ingestible wireless capsule for the visualisation of the small intestine mucosa. The device has received approval from Food and Drug Administration and has gained momentum since it has been more successful than traditional techniques in the diagnosis of small intestine disorders. In 2004 an esophagus specific capsule was launched, while a solution for colon is still under development. However, present solutions suffer from several limitations: they move passively by exploiting peristalsis, are not able to stop intentionally for a prolonged diagnosis, they receive power from an internal battery with short length, and their usage is restricted to one organ, either small bowel or esophagus. However the steady progresses in many branches of engineering, including microelectromechanical systems (MEMS), are envisaged to affect the performances of capsular endoscopy. The near future foreshadows capsules able to pass actively through the whole gastrointestinal tract, to retrieve views from all organs and to perform drug delivery and tissue sampling. In the long term, the advent of robotics could lead to autonomous medical platforms, equipped with the most advanced solutions in terms of MEMS for therapy and diagnosis of the digestive tract. In this review, we discuss the state of the art of wireless capsule endoscopy (WCE): after a description on the current status, we present the most promising solutions.

Capsule Endoscopes↗

In vivo robotic capsules: determination of the number of turns of its power receiving coil.

This paper introduces a strategy how to determine the number of turns of the receiving coil for in vivo robotic capsules to maximize the regulated load power. For the expected functionality additional to the endoscopic function, it is necessary to employ wireless power transmission and increase the receiving power to a maximum. The functional relation between the number of turns and the power is derived and verified by implementing a ferrite-loaded receiver coil.

Capsule Endoscopes↗

Magnetic shells for gastrointestinal endoscopic capsules as a means to control their motion.

Wireless endoscopic video capsules are becoming largely used today for non-invasive explorations of the digestive tube. Traditional types of such capsules present a major drawback: their motion can not be controlled, since they proceed by means of the visceral peristalsis only. In order to enable a wireless control of their motion, this paper describes a new concept based on the use of external magnetic fields. The proposed technique exploits magnetic shells to be applied to traditional capsules prior to their use. The shells are capable of interacting with an imposed external magnetic field, providing a means to control the capsule movement and orientation. This solution is readily and cheaply applicable to any commercial endoscopic capsule, avoiding internal modifications. The paper reports results of preliminary bench tests. Prototype elastic shells made of a silicone elastomer mixed with magnetic particles were fabricated and tested with the Given Imaging M2A capsule in simplified experimental conditions. These tests permitted to demonstrate translations, rotations and roto-translations of the capsule/shell complex onto pieces of bovine tissues.