We present a virtual reality simulator to realize interventional radiology (IR) procedures remotely. The simulator contains two subsystems: one at the local site and the other at the remote site. At the local site, the interventional radiologist interacts with a three-dimensional (3-D) vascular model extracted from the patient's data and inserts IR devices through the Motion Tracking Box (MTB), which converts physical motion (translation and rotation) of IR devices into digital signal. This signal is transferred to the Actuator Box (AB) at the remote site that drives the IR devices in the patient. The status of the IR devices is subsequently fed back to the local site and displayed on the vascular model. To prove the concept, the prototype developed employs a physical angiography phantom (mimicking the patient) and its corresponding 3-D digital model. A magnetic tracking system provides information about positioning of the IR devices in the phantom. The initial results are encouraging. The AB controlled remotely drives IR devices with resolution of 0.00288 mm/step in translation and 0.079 deg/step in rotation.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Ma Xin, Zhao Lei, Ihar Volkau, Zheng Weili, Aamer Aziz, Marcelo H Ang, Wieslaw L Nowinski. 2006. A virtual reality simulator for remote interventional radiology: concept and prototype design.. https://doi.org/10.1109/tbme.2006.873762

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Transhepatic vascular access for diagnostic and interventional procedures: techniques, outcome, and complications.

Early or multiple cardiac catheterizations may result in occlusion of the femoral veins prohibiting their use. The internal jugular or sub-clavian approach may be an appropriate alternative. However, these approaches may not be suitable in patients with surgical interruption of the superior vena cava. In other patients, they may not allow easy access to certain areas of the heart. The transhepatic approach is an important alternative route for performing cardiac catheterization and interventions. Depending on the planned procedure, it may be the preferred route to perform the cardiac catheterization even in the presence of patent femoral veins. The indications, technical details and potential complications of this approach are discussed.

Angiography↗

Embolization for a bleeding pelvic fracture in a patient with persistent sciatic artery.

Emergency embolization for a bleeding pelvic fracture was performed in a patient with persistent sciatic artery, a rare anatomic variation in which the internal iliac artery continues to the popliteal artery. Successful hemostasis was achieved without complications by means of selective catheterization into each branch of the internal iliac artery. Awareness of this unusual vascular anatomy is critically important to avoid serious complication of emergency embolization.

Angiography↗