PubMed · 16958057
Imaging artifacts at 3.0T.
Abstract
Clinical MRI at a field strength of 3.0T is finding increasing use. However, along with the advantages of 3.0T, such as increased SNR, there can be drawbacks, including increased levels of imaging artifacts. Although every imaging artifact observed at 3.0T can also be present at 1.5T, the intensity level is often higher at 3.0T and thus the artifact is more objectionable. This review describes some of the imaging artifacts that are commonly observed with 3.0T imaging, and their root causes. When possible, countermeasures that reduce the artifact level are described.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Matt A Bernstein, John Huston, Heidi A Ward. 2006. Imaging artifacts at 3.0T.. https://doi.org/10.1002/jmri.20698. https://pubmed.health/record/pubmed%3A16958057/imaging-artifacts-at-3-0t