PubMed · 17171406
The development of a multichannel electrode array for retinal prostheses.
Abstract
The development of a multielectrode array is the key issue for retinal prostheses. We developed a 10 x 10 platinum electrode array that consists of an 8-microm polyimide layer sandwiched between 5-microm polymonochloro-para-xylylene (parylene-C) layers. Each electrode was formed as a 30-microm-high bump by Pt/Au double-layer electroplating. We estimated the charge delivery capability (CDC) of the electrode by measuring the CDCs of two-channel electrode arrays. The dimensions of each electrode of the two-channel array were the same as those of each electrode formed on the 10 x 10 array. The results suggest that for cathodic-first (CF) pulses, 80% of electrodes surpassed our development target of 318 microC/cm2, which corresponds to the charge density of pulses of 500 micros duration and 200 microA amplitude for a 200-microm-diameter planar electrode.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Yasuo Terasawa, Hiroyuki Tashiro, Akihiro Uehara, Tohru Saitoh, Motoki Ozawa, Takashi Tokuda, Jun Ohta. 2006-12-21. The development of a multichannel electrode array for retinal prostheses.. https://doi.org/10.1007/s10047-006-0352-1
Cite the original work for its findings. Save a collection to share your selection of sources.