Abstract

BACKGROUND: Although few studies have been performed recently, several have suggested that some practitioners are not well able to detect preset anesthesia machine faults. METHODS: We performed a prospective study to determine whether there is a correlation between duration of anesthesia practice and the ability to detect anesthesia machine faults. Our hypothesis was that more anesthesia practice would increase the ability to detect anesthesia machine faults. This study was performed during a nationally attended anesthesia meeting held at a large academic medical center, where 87 anesthesia providers were observed performing anesthesia machine checkouts. The participants were asked to individually check out an anesthesia machine with an unspecified number of preset faults. The primary outcome measures were the written listing of faults detected during an anesthesia machine checkout. RESULTS: Of the five faults preset into the test machine, participants with 0-2 yr experience detected a mean of 3.7 faults, participants with 2-7 yr experience detected a mean of 3.6 faults, and participants with more than 7 yr experience detected a mean of 2.3 faults (P < 0.001). CONCLUSIONS: Our prospective study demonstrated that anesthesia machine checkout continues to be a problem.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Eric R Larson, Gregory A Nuttall, Brian D Ogren, Dean D Severson, Sarah A Wood, Laurence C Torsher, William C Oliver, Mary E Shirk Marienau. 2007. A prospective study on anesthesia machine fault identification.. https://doi.org/10.1213/01.ane.0000250225.96165.4b

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Perception of training needs and opportunities in advanced airway skills: a survey of British and Irish trainees.

BACKGROUND AND OBJECTIVE: We surveyed delegates at the Group of Anaesthetists in Training (UK) meeting to investigate evidence of a training-gap (number of fibreoptic intubations believed to bestow competence vs. number actually performed). METHODS: Questionnaires were distributed to and collected from delegates in person. Questions covered six areas, including experience of fibreoptic intubation and cricothyrotomy, fibreoptic intubation as a specialist skill and ethical issues. RESULTS: We received 221 replies (76%). All trainees believed competence to be achievable with 10 intubations (interquartile range (IQR) 10-20); the median number performed was 2 (IQR 0-4). This was statistically significant for the groups' senior house officers, 1st and 2nd year registrars and 3rd and 4th year registrars; P < 0.0001. Many final year trainees (12/20, 60%) also failed to achieve their competency target. Few trainees had seen or performed any cricothyrotomies (medians 0, IQRs 0-1 and 0-0). Most (195/208, 94%) believed that fibreoptic intubation was a core skill and 199/212 (94%) believed that all should be competent by completion of training. Ten percent (n = 208) felt it unethical to perform an awake training intubation with full consent and 10% believed it acceptable without explanation. Most (82.7%) would fibreoptically intubate an asleep patient (requiring intubation) without consent. CONCLUSION: Trainees reported a gap between their perception of competence and achievement in awake fibreoptic intubation. Simple and complex simulations and structured training programmes may help. Anaesthetists must address the ethics of clinical training in advanced airway management.

Anesthesiology↗