Hybrid multidirectional test method to evaluate spinal adjacent-level effects.

Abstract

BACKGROUND: Several clinical studies have documented long-term adjacent-level effects of spinal fusion, due to stress concentration and motion loss at the fused segment. Non-fusion motion preservation devices are designed to eliminate or slow down such adverse effects. Therefore, appropriate biomechanical evaluation of the adjacent-level effects in spine is important and timely. Although many biomechanical studies are available and have provided some understanding of the adjacent-level effects, results have large variation and are conflicting, mostly due to the use of inappropriate and ill-defined methods. A new test method especially designed to study spinal adjacent-level effects is needed. METHODS: The proposed Hybrid method uses unconstrained pure moment to provide rotation-input for multi-directional testing. The new method has four steps: (1) Intact spine specimen with entire mobile region is used. The specimen is prepared to measure various biomechanical parameters, e.g., disc pressures, ligament strains, and facet loads. (2) Appropriate unconstrained pure moment is applied to the intact specimen and total range of motion is determined. (3) Unconstrained pure moment is applied to the spinal construct (specimen with an implant) until the total range of motion of the construct equals that of the intact. (4) Statistical comparison of the biomechanical parameters between the construct and intact quantifies the adjacent-level effects. FINDINGS: The uniqueness of the proposed method, to study the adjacent level effects due to fusion and non-fusion devices, is that it applies the needed rotation-input to the spine specimen, using available methodology with minimal modification. INTERPRETATION: Previous studies have lacked appropriate and well-defined methodologies to evaluate spinal adjacent-level effects. The proposed method uses well-known methodology and yields high quality, and laboratory-independent results for the fusion and non-fusion devices.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Manohar M Panjabi. 2006-12-29. Hybrid multidirectional test method to evaluate spinal adjacent-level effects.. https://doi.org/10.1016/j.clinbiomech.2006.08.006

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

The effect of excessive tibial torsion on the capacity of muscles to extend the hip and knee during single-limb stance.

Excessive tibial torsion, a rotational deformity about the long axis of the tibia, is common in patients with cerebral palsy who walk with a crouch gait. Previous research suggests that this deformity may contribute to crouch gait by reducing the capacity of soleus to extend the knee; however, the effects of excess external torsion on the capacity of other muscles to extend the stance limb during walking are unknown. A computer model of the musculoskeletal system was developed to simulate a range of tibial torsion deformities. A dynamic analysis was then performed to determine the effect of these deformities on the capacity of lower limb muscles to extend the hip and knee at body positions corresponding to the single-limb stance phase of a normal gait cycle. Analysis of the model confirmed that excessive external torsion reduces the extension capacity of soleus. In addition, our analysis revealed that several important muscles crossing the hip and knee are also adversely affected by excessive tibial torsion. With a tibial torsion deformity of 30 degrees , the capacities of soleus, posterior gluteus medius, and gluteus maximus to extend both the hip and knee were all reduced by over 10%. Since a tibial torsion deformity reduces the capacity of muscles to extend the hip and knee, it may be a significant contributor to crouch gait, especially when greater than 30 degrees from normal, and thus should be considered by clinicians when making treatment decisions.

Biomechanical Phenomena↗

Predicting the subject-specific primary stability of cementless implants during pre-operative planning: preliminary validation of subject-specific finite-element models.

Pre-operative planning help the surgeon in taking the proper clinical decision. The ultimate goal of this work is to develop numerical models that allow the surgeon to estimate the primary stability during the pre-operative planning session. The present study was aimed to validate finite-element (FE) models accounting for patient and prosthetic size and position as planned by the surgeon. For this purpose, the FE model of a cadaveric femur was generated starting from the CT scan and the anatomical position of a cementless stem derived by a skilled surgeon using a pre-operative CT-based planning simulation software. In-vitro experimental measurements were used as benchmark problem to validate the bone-implant relative micromotions predicted by the patient-specific FE model. A maximum torque in internal rotation of 11.4 Nm was applied to the proximal part of the hip stem. The error on the maximum predicted micromotion was 12% of the peak micromotion measured experimentally. The average error over the entire range of applied torques was only 7% of peak measurement. Hence, the present study confirms that it is possible to accurately predict the level of primary stability achieved for cementless stems using numerical models that account for patient specificity and surgical variability.

Biomechanical Phenomena↗

Experimental validation of intact and implanted distal femur finite element models.

Four finite element (FE) models of intact and distal femur of knee replacements were validated relative to measured bone strains. FE models of linear tetrahedrons were used. Femoral replacements with cemented stemless, cemented and noncemented femoral stems of the PFC Sigma Modular Knee System were analyzed. Bone strains were recorded at ten locations on the cortex. The magnitude of the FE bone strains corresponded to the mean measured strains, with an overall agreement of 10%. Linear regression between the FE and mean experimental strains produced slopes between 0.94 and 1.06 and R(2) values between 0.92 and 0.99. RSME values were less than 12%. The FE models were able to adequately replicate the mechanical behavior of distal femur reconstructions.

Biomechanical Phenomena↗