PubMed · 1732459
Thyroid uptake neck phantoms are not created equal.
Abstract
The main purpose of this communication is to alert nuclear medicine departments to the fact that the earlier version of the water phantom grossly overestimates soft-tissue attenuation in the neck, resulting in calculated thyroid uptake values which are significantly overestimated (in hyperthyroid patients we noted uptake values approaching or exceeding 100%). We believe that the solid Lucite phantom (which is the one recommended by IAEA) better approximates the human neck soft tissue overlying the thyroid. Institutions that continue to use the water phantom should be aware that their thyroid uptakes will be relatively elevated and the normal range must be shifted accordingly. Our normal range is 10%-30% uptake of 123I at 24 hr for the Lucite phantom. For the water phantom, the estimated normal range would be 15%-45%. In addition, the phantom type should be considered when comparing uptake results with those from another institution for a particular patient. Also, treatment doses for Grave's disease could be significantly affected, if such doses are calculated by a formula that depends on uptake.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
G A Vahjen, R C Lange, T F Merola. 1992. Thyroid uptake neck phantoms are not created equal.. https://pubmed.ncbi.nlm.nih.gov/1732459/
Cite the original work for its findings. Save a collection to share your selection of sources.