A total body nitrogen facility for paediatric use.
Abstract
The design, calibration and evaluation of a facility for in vivo prompt gamma neutron capture analysis of total body nitrogen in children is described. The patient is scanned in both supine and prone positions, across a vertically collimated beam from a 1 GBq 252Cf fission source. Two NaI(T1) detectors are placed on either side of the patient, perpendicular to both the neutron beam and the scanning direction. The effective dose equivalent delivered to a child during an 840 s scan is approximately 0.14 mSv (QF = 10). Correction factors for nitrogen background (width-dependent), hydrogen background (1-4% of hydrogen gamma ray peak) and the differential attenuation of nitrogen and hydrogen gamma rays (width-dependent) can be applied to the measured nitrogen-to-hydrogen gamma ray counts ratio. By using the mass of hydrogen (based on body mass and fat mass) as an internal standard, the nitrogen mass can then be determined. Measurements with a urea-containing box phantom show that the current precision (CV) of the net nitrogen counts and of the nitrogen-to-hydrogen counts ratio is +/- 2.0% and +/- 1.5%, respectively. Using small anthropomorphic and other phantoms, estimation of the mass of nitrogen has a precision of +/- 1.4 to 5.4% and an accuracy of 97.1 to 101.5%.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
L A Baur, B J Allen, A Rose, N Blagojevic, K J Gaskin. 1991. A total body nitrogen facility for paediatric use.. https://doi.org/10.1088/0031-9155%2F36%2F10%2F005