PubMed · 1795035
Increasing thermospray response for cortisol by derivatization.
Abstract
The 21-hydroxyl group of cortisol was selectively acetylated under mild conditions without affecting the more sterically hindered 11 beta- and 17 alpha-hydroxyl groups. The reaction was performed with a mixture of acetic anhydride and triethylamine in acetonitrile and was complete in less than 15 min at room temperature. The thermospray mass spectrum of cortisol 21-acetate showed minimal fragmentation with the [M + H]+ ion as the base peak. High sensitivity was achieved for acetylated cortisol during selected ion monitoring, the signal-to-noise ratio being increased by a factor of about 4, compared to underivatized cortisol. The limit of detection of cortisol 21-acetate was estimated at 0.24 pmol injected, making thermospray liquid chromatography-mass spectrometry competitive with gas chromatography-mass spectrometry for the determination of cortisol in biological fluids.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
J Paulson, C Lindberg. 1991-08-21. Increasing thermospray response for cortisol by derivatization.. https://doi.org/10.1016/s0021-9673(01)88445-4
Cite the original work for its findings. Save a collection to share your selection of sources.