Coated amperometric electrode arrays for multicomponent analysis.

Abstract

This paper describes a sensor array of several amperometric electrodes, each coated with a different permselective film. Coatings with different transport properties, based on size (cellulose acetate), charge (Nafion, poly(vinylpyridine), poly(ester sulfonic acid], and polarity (phospholipid) are employed in connection with a four-electrode thin-layer flow detector. With equipotential operation, the array's response pattern of each analyte provides a unique characterization of the individual components. Multicomponent analysis is obtained by taking advantage of the partial selectivity of the individual sensors and using a pattern recognition (multiple linear regression) method. Additional information is obtained by recording the complete hydrodynamic voltammogram at the individual sensors. The merits of the new array are illustrated for the quantification of neurologically significant catechol compounds.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

J Wang, G D Rayson, Z L Lu, H Wu. 1990-09-15. Coated amperometric electrode arrays for multicomponent analysis.. https://doi.org/10.1021/ac00217a006

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Effect of aging the hollow cathode by sputtering on the analytical precision of the hollow cathode discharge emission source.

Sputtering changes the analytical performance of the hollow cathode as an emission source. Proper conditioning (aging) of the hollow with sputtering greatly improves precision. Conditioning decreases the blank emission signal and changes the shape of the hollow. Micrographs which show the evolution of the bottom of the hollow into a stable bulb shape are presented. The blank emission signal decreases with aging of the hollow and approaches a constant value when the hollow is properly conditioned. A short-term stability (n = 5) of approximately 1.8% RSD has been observed frequently. An average long-term stability over several days (n = 45) of 4.5% for Li and 3.5% for Na in the emission signal from microsamples (less than 50 nL) deposited in the hollow cathode discharge source is reported here. Instrumentation, operation, and sample preparation procedures are described.

Electrodes