Energy constancy checking for electron beams using a wedge-shaped solid phantom combined with a beam profile scanner.

Abstract

An energy constancy checking method is presented which involves a specially designed wedge-shaped solid phantom in combination with a multiple channel ionization chamber array known as the Thebes device. Once the phantom/beam scanner combination is set up, measurements for all electron energies can be made and evaluated without re-entering the treatment room. This is also valid for the readjustment of beam energies which are found to deviate from required settings. The immediate presentation of the measurements is in the form of crossplots which resemble depth dose profiles. The evaluation of the measured data can be performed using a hand-held calculator, but processing of the measured signals through a PC-type computer is advisable. The method is insensitive to usual fluctuations in beam flatness. The sensitivity and reproducibility of the method are more than adequate. The method may also be used in modified form for photon beams.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

U F Rosenow, M K Islam, H Gaballa, H Rashid. Energy constancy checking for electron beams using a wedge-shaped solid phantom combined with a beam profile scanner.. https://doi.org/10.1118/1.596719

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Technical note: reconstructing dose distributions from manually planned electron boosts in breast radiotherapy.

PURPOSE: In breast radiotherapy, delivery of manually-calculated electron boosts limits retrospective dose-response analyses as dose distribution is unavailable. This work evaluates the feasibility of reconstructing dose distributions from manually planned electron boosts in breast-conserving radiotherapy. METHODS: Only 72 out of 198 breast cancer patients had complete stored dose distributions from sequential electron boosts in the REQUITE study. Arbitrary data from 70/72 patients were used to develop and validate dose reconstruction method. Twenty patients were used to determine optimal parameters for Monte-Carlo-based (MC) electron dose reconstruction on RayStation (v.11B-R), considering CT-calibration curve, MC-history number, andcalculation grid resolution. Remaining 50 patients were used to quantify dose reconstruction accuracy. The similarity between reconstructed and stored dose was evaluated using 3D-gamma index and dosimetric parameters extracted from breast and tumour bed contours. Dose difference location was evaluated using dose-location histogram. RESULTS: Calculation grid resolution significantly impacted electron dose distribution (p&#xa0;<&#xa0;0.01), where the finest grid (0.15&#xa0;cm) showed highest similarity to stored doses. CT-calibration curve and MC-history number had a negligible influence on dose reconstruction. Dosimetric difference between reconstructed and stored doses was&#xa0;<&#xa0;1&#xa0;Gy for breast and tumour bed. Reconstructed dose was achieved&#xa0;>&#xa0;90% gamma passing rate in the validation set. However, around 2.5&#xa0;Gy dose differences were observed at the skin and tissue interface regions. CONCLUSIONS: Retrospective electron boost dose reconstruction is feasible with acceptable accuracy, and could increase data completeness in large cohort studies. Caution is advised when assessing dose near tissue interface and further validation is needed outside the REQUITE dataset.

Electrons

Application of mass and energy balance regularities in fermentation. Reprinted from Biotechnology and Bioengineering, Vol. XX, No. 10, Pages 1595-1621 (1978).

Material and energy balances for fermentation processes are developed based on the facts that the heat of reaction per electron transferred to oxygen for a wide variety of organic molecules, the number of available electrons per carbon atom in biomass, and the weight fraction carbon in biomass are relatively constant. Mass-energy balance equations are developed which relate the biomass energetic yield coefficient to sets of variables which may be determined experimentally. Organic substrate consumption, biomass production, oxygen consumption, carbon dioxide production, heat evolution, and nitrogen consumption are considered as measured variables. Application of the balances using direct and indirect methods of yield coefficient estimation is illustrated using experimental results from the literature. Product formation is included in the balance equations and the effect of product formation on biomass yield estimates is examined. Application of mass-energy balances in the optimal operation of continuous single-cell protein production facilities is examined, and the variation of optimal operating conditions with changes in yield are illustrated for methanol as organic substrate.

Electrons

Use of an electron reflector to improve dose uniformity at the vertex during total skin electron therapy.

PURPOSE: The vertex of the scalp is always tangentially irradiated during total skin electron therapy (TSET). This study was conducted to determine the dose distribution at the vertex for a commonly used irradiation technique and to evaluate the use of an electron reflector, positioned above the head, as a means of improving the dose uniformity. METHODS AND MATERIALS: Phantoms, simulating the head of a patient, were irradiated using our standard procedure for TSET. The technique is a six-field irradiation using dual angled electron beams at a treatment distance of 3.6 meters. Vertex dosimetry was performed using ionization methods and film. Measurements were made for an unmodified 6 MeV electron beam and for a 4 MeV beam obtained by placing an acrylic scattering plate in the beam line. Studies were performed to examine the effect of electron scattering on vertex dose when a lead reflector, 50 x 50 cm in area, was positioned above the phantom. RESULTS: The surface dose at the vertex, in the absence of the reflector, was found to be less than 40% of the prescribed skin dose. Use of the lead reflector increased this value to 73% for the 6 MeV beam and 99% for the degraded 4 MeV beam. Significant improvements in depth dose were also observed. The dose enhancement is not strongly dependent on reflector distance or angulation since the reflector acts as a large source of broadly scattered electrons. CONCLUSION: The vertex may be significantly underdosed using standard techniques for total skin electron therapy. Use of an electron reflector improves the dose uniformity at the vertex and may reduce or eliminate the need for supplemental irradiation.

Electrons