Specificity of two-site immunometric assays.
Abstract
The assessment of specificity in two-site immunometric assays is complex. Conventional approaches based on standard substitution with cross-reactant may yield misleading results. An 'in-house' immunoradiometric assay (IRMA) for human growth hormone (hGH) appeared free from cross-reactivity with human placental lactogen (hPL) in a single point standard substitution experiment performed by the United Kingdom External Quality Assessment Scheme (UK EQAS). However, detailed studies based on the substitution of hGH standards with hPL cross-reactant, the addition of hPL cross-reactant to hGH standards and the recovery of added hGH from endogenous cross-reactant (sera from pregnant subjects) revealed gross cross-reactivity of hPL in the hGH assay. We recommend that the validation of assay specificity should require the demonstration of quantitative recovery of a range of analyte concentrations from human sera that contain varying endogenous concentrations of cross-reactant.