PubMed · 2299229
Development of inductive heating equipment using an inductive aperture-type applicator.
Abstract
We designed an inductive aperture-type applicator, IATA, consisting of a one-turn, square, column-like coil made of a metal strip. In experiments using a phantom, this applicator was seen to have the following characteristics: (1) it does not excessively heat the fat layer; (2) deep portions can be heated effectively; (3) heat can be applied to the subject, without contact with the applicator; (4) excessive generation of heat does not occur in parts of tissue near the edge of the aperture; (5) size of the aperture can be changed for adaptation to size of the site to be heated; (6) the aperture can be modified to fit the shape of the subject and the depth of penetration is influenced by size of the subject being heated; (7) the applicator can be driven at an arbitrary frequency; (8) efficiency of heat generation to the input power is low. The inductive heating equipment, 'Inductron', which we have developed possesses the necessary requirements for adequate clinical application.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
H Kato, M Furukawa, N Uchida, T Kasai, Y Fujita, F Koda, H Kuroda, T Ishida. Development of inductive heating equipment using an inductive aperture-type applicator.. https://doi.org/10.3109/02656739009140812