Gamma camera image acquisition, display, and processing with the personal microcomputer.

Abstract

The authors evaluated the potential of a microcomputer for direct acquisition, display, and processing of gamma camera images. Boards for analog-to-digital conversion and image zooming were designed, constructed, and interfaced to the Macintosh II (Apple Computer, Cupertino, Calif). Software was written for processing of single, gated, and time series images. The system was connected to gamma cameras, and its performance was compared with that of dedicated nuclear medicine computers. Data could be acquired from gamma cameras at rates exceeding 200,000 counts per second, with spatial resolution exceeding intrinsic camera resolution. Clinical analysis could be rapidly performed. This system performed better than most dedicated nuclear medicine computers with respect to speed of data acquisition and spatial resolution of images while maintaining full compatibility with the standard image display, hard-copy, and networking formats. It could replace such dedicated systems in the near future as software is refined.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

J L Lear, J P Pratt, D R Roberts, T Johnson, A Feyerabend. 1990. Gamma camera image acquisition, display, and processing with the personal microcomputer.. https://doi.org/10.1148/radiology.175.1.2315488

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Implications of dual-energy-window (DEW) scatter correction inaccuracies for 111In quantitative geometric mean imaging.

There is increasing clinical interest in the use of quantitative imaging for radiopharmaceuticals labelled with 111In. Dual-energy-window (DEW) scatter correction is a frequently used component of planar geometric mean quantitative imaging, but it is known that the scatter multiplier k suffers from significant dependence on the characteristics of the scatter medium. Phantom studies with a variety of source geometries were carried out to determine the clinical impact of this dependence on the quantitative accuracy of tumour imaging carried out in conjunction with attenuation correction. Spheres of various sizes (5-20 ml volumes) containing approximately 3.7 MBq (100 microCi) 111In were imaged at a variety of depths (4.8-10.5 cm) within an elliptical water-filled phantom, as well as in air. Geometric mean emission images were acquired using a 20% photopeak window at 247 keV and a 10% scatter window at 205 keV. These emission images were corrected for attenuation using measured 99Tcm transmission data that were scaled to 111In photon energies. Scatter correction was performed in two ways: (1) using the standard DEW method and (2) using a modified DEW method that takes into account benign scatter in the detector crystal. Errors in the activity estimates ranged from -4% to +3% for method 1 in water, and -5% to +3% for method 2 in water. In air, method 1 ranged from -13% to -5%, and method 2 ranged from -10% to -1%. Method 1 was found to yield an accuracy equivalent to that of method 2, except in conditions of very low patient scatter, when the modified method behaved significantly better. We conclude that in a variety of realistic geometries, variations in scatter fraction as determined by the DEW scatter correction method combined with appropriate attenuation correction need not inhibit accurate absolute quantitation of spherical 'tumours' labelled with 111In when using planar imaging.

Gamma Cameras

[Energy spectral analysis in a photopeak region of 201Hg X-rays for 201Tl imaging].

For thallium-201 (201Tl) imaging, a photopeak of X-ray region of mercury-201 (201Hg) around 70 keV is usually employed. In such a low energy region, scatter and attenuation in the body become serious problems for quantitative scintigraphic imaging. We measured energy spectra around 201Hg X-ray region by a 201Tl line source with scattering materials and a gamma camera, and analyzed them to investigate the amount of scattered photons in 201Tl imaging window. Furthermore, Monte Carlo calculations were performed to distinguish primary photons from scattered photons. The results of analysis of energy spectra showed that the 201Hg X-ray photopeak consists of four different X-rays and do not have a perfect Gaussian shape. It was also found that not only forward scatter but also much backward scatter are included in this region. These scattered photons were reduced by using an off-peak window moved toward the higher energy region. We conclude that the data obtained from the analysis of scattered photons in 201Tl imaging window are very useful to develop a new scatter correction method and evaluate the scatter rejection methods proposed.

Gamma Cameras