The electrogastric signal can be measured cutaneously on the abdomen. This is attractive because it is harmless to patients or volunteers. However, the poor quality of the cutaneous measurements necessitates signal enhancements. Hence, in this paper, an adaptive multichannel signal enhancing system is proposed. The mu-vector least mean square (LMS) algorithm is applied to adjust the weights of the adaptive filters in the system. The detailed description and the performance analysis of the system is given in the paper. Applying the proposed system, the respiratory artifact, the electrode-skin noise, some of motion artifacts, and the electrocardiography (ECG) can be efficiently reduced while the characteristics of the relevant gastric signal is less affected.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
J D Chen, J Vandewalle, W Sansen, G Vantrappen, J Janssens. 1990. Multichannel adaptive enhancement of the electrogastrogram.. https://doi.org/10.1109/10.52329