Examination of explanted polyurethane pacemaker leads using the scanning electron microscope.

Abstract

Since 1978, 2,365 polyurethane (PU) insulated cardiac pacing leads were implanted transvenously at our institution. To date, there have been no insulation failures in those leads. Thirty-seven PU leads were explanted, mainly for exit block, and 28 of these were investigated using the scanning electron microscope. We found a homogeneous distribution of surface changes in all lead segments in 56% of the 28 examined. These changes were more pronounced at the ligature site; severe surface cracking was noticed in 21%, with the deepest crack being 40 micron (average range of 10-15 micron). There appeared to be no time-dependency of the surface changes as indicated by regression analysis (r = 0.32, p greater than 0.05). The ultimate severity and outcome of this degradation process in the leads reported in this study will only be known in the future after longer use. We conclude that excess stress must be avoided during the implantation procedure and that careful surveillance is necessary.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

F Beyersdorf, J Kreuzer, L Schmidts, P Satter. 1985. Examination of explanted polyurethane pacemaker leads using the scanning electron microscope.. https://doi.org/10.1111/j.1540-8159.1985.tb05860.x

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Tip cooling effect and failure mechanism of field-emitting carbon nanotubes.

The cooling effect accompanying field electron emission has been considered for a single carbon nanotube (CNT) used as a field emission (FE) electron source. An improved model for the failure mechanism of field emitting CNTs has been proposed and validated. Our model predicts a maximum temperature (T-max) located at an interior point rather than the tip of the CNTs, and the failure of the CNT emitters tends to take place at the T-max point, inducing a segment by segment breakdown process. A combination of Joule heating and electrostatic force effect is proposed responsible for initiating the failure of the field emitting CNT and validated by in situ FE observation.

Equipment Failure↗

[Importance of an initial check-up for diagnostic radiology apparatus management].

The management of apparatus for diagnostic imaging is an important job for radiological technologists. In the last 10 years we have encountered 2227 malfunctions in 3652 days. We detected 163 cases at the initial check-up, which accounted for 7.3% of total cases. By performing an initial check-up, we detected one malfunction every 21 days and were able to keep the effects of mishaps to a minimum, prevent accidents, and obtain stable-quality images.

Equipment Failure↗