Abstract

Acute and chronic threshold measurements were compared for three different electrodes: the Medtronic Target Tip model 4011, the Medtronic Ring Tip Model 6971 and the Siemens-Elema Carbon Tip Model 412S. The results of this study show that acute thresholds are lower with the Target Tip lead as compared to the other two. This finding applies to chronic thresholds as well. The results of this study favor programming to lower output energies while maintaining adequate safety margins with this lead system.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

M Djordjevic, P Stojanov, D Velimirovic, D Kocovic. 1986. Target lead--low threshold electrode.. https://doi.org/10.1111/j.1540-8159.1986.tb06695.x

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Visual and quantitative ictal EEG predictors of outcome after temporal lobectomy.

PURPOSE: We investigated whether visual and quantitative ictal EEG analysis could predict surgical outcome after anteromesial temporal lobectomy (AMTL) in which mesial structures, basal, and temporal tip cortex were resected. METHODS: We retrospectively reviewed 282 presurgical scalp-recorded ictal EEGs (21- to 27-channel) from 75 patients who underwent AMTL. We examined the pattern of seizure onset (frequency, distribution, and evolution) and estimated the principal underlying cerebral generators by using a multiple fixed dipole model that decomposes temporal lobe activity into four sublobar sources (Focus 1.1). We correlated findings with a 2-year postoperative outcome. RESULTS: Sixteen patients had seizures with a well-lateralized, regular 5 to 9-Hz rhythm at onset, that most often had a temporal or subtemporal distribution. All patients became seizure free after surgery. In 51 patients, seizure onset was remarkable for lateralized slow rhythms (<5 Hz), which sometimes appeared as periodic discharges, were often irregular and stable only for short periods (<5 s), and had a widespread lateral temporal distribution. Among these a favorable surgical outcome was encountered in patients with seizures having prominent anterior-tip sources ( 16 of 17 seizure free), whereas those with dominant lateral or oblique sources had a less favorable outcome (three of 14 and 13 of 18, respectively). Irregular, nonlateralized slowing characterized seizure onsets in eight patients. Three patients became seizure free after surgery. CONCLUSIONS: Both visual and quantitative sublobar source analysis of scalp ictal EEG can predict surgical outcome in most cases after AMTL and complement non-invasive presurgical evaluation.

Electrodes, Implanted