Severe dialyzer dysfunction undetectable by standard reprocessing validation tests.

Abstract

It is generally accepted that careful monitoring of total cell volume and ultrafiltration rates will ensure adequate function of reprocessed dialyzers. During routine urea kinetic measurements we noted that the percent of patients with clearances less than 200 ml/min increased from 5% to 48% despite adherence to these validation tests. As these patients did not have evidence of recirculation in the vascular access, possible causes of dialyzer dysfunction were investigated. Injection of methylene blue into the dialysate port revealed non-uniform flow of dialysate in dialyzers from patients with markedly reduced clearances. In vitro studies of dialyzers subjected to sequential daily reprocessing, without patient exposure, demonstrated that in vitro clearances declined in one lot but not another. The initial clearances of 218 +/- 4 ml/min fell progressively to 112 +/- 18 (P less than 0.001) after 15 reuses. No effects of reprocessing were found in a different lot (230 +/- 2 vs. 226 +/- 4 ml/min). Soaking the dialyzers from the affected lot in either the disinfectant or dialysate solution caused a decline in the clearances which was less than that of serial reuse. Although the magnitude of the problem of dialyzer malfunction with reuse is unknown, careful attention to dialyzer function is warranted in patients treated with reprocessed dialyzers.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

J A Delmez, C A Weerts, P D Hasamear, D W Windus. 1989. Severe dialyzer dysfunction undetectable by standard reprocessing validation tests.. https://doi.org/10.1038/ki.1989.220

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Screening for neuroblastoma in the north of England.

OBJECTIVE: To determine the feasibility of establishing a system of screening for neuroblastoma. DESIGN: Prospective study of mass screening in four clearly defined geographical areas. SETTING: Four health districts of the Northern region of England. SUBJECTS: 20,829 babies aged 6 months, 92% of target population. INTERVENTIONS: Collection of urine on filter paper for analysis of content of homovanillic and vanillylmandelic acid in relation to urinary creatinine concentrations. MAIN OUTCOME MEASURES: Derivation of reference range. Identification of babies with homovanillic or vanillylmandelic acid > 3 SD above the mean (positive cases). Investigation of positive cases for evidence of neuroblastoma. RESULTS: The upper limit of normal (3 SD above the mean) for vanillylmandelic acid was 15 mumol/mmol creatinine and for homovanillic acid 24 mumol/mmol creatinine. Of the 20,829 babies screened, 2537 (12.2%) required a second sample to be taken because the first sample was inadequate. Of these, 527 (2.5%) provided a liquid urine specimen and 10 (0.04%) had positive results for neuroblastoma. Two of them had neuroblastoma (true positives) and eight did not (false positives). A further three children from the cohort were subsequently found to have neuroblastoma; they had raised homovanillic acid or vanillylmandelic acid values, or both, but screened negative at 6 months. CONCLUSIONS: Screening for neuroblastoma is possible in the health care system of the United Kingdom. Evaluation of the efficacy of screening in reducing the mortality from neuroblastoma requires a controlled trial.

Creatinine

More on SI units.

Explore the source record for details and available documents.

Creatinine