In the present studies we have evaluated the optimal operating conditions for the Hamilton-Thorn HTM-2000 computerized semen analyzer (Hamilton-Thorn, Danvers, MA). The best reproducibility in measurement of sperm concentration was obtained using 20 frames acquired at 19 frames/s. The measurement of sperm concentration was not adversely affected by the number of fields analyzed. The intrasample and intersample coefficients of variation for sperm concentration were 9.5% and 25.5%; sperm motility, 18.4% and 28.9%; lateral head displacement, 16.5% and 19.9%; path velocity, 6.8% and 13.9%; progressive velocity, 4.5% and 9.9%; and linear index, 2.5% and 4.2%; respectively. These differences suggest that sampling error has a significant influence on the reliability of sperm evaluation. The precision and rapidity of the HTM-2000 compares favorably with data previously reported from other systems available for clinical semen analysis.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
N G Pedigo, M W Vernon, T E Curry. 1989. Characterization of a computerized semen analysis system.. https://doi.org/10.1016/s0015-0282(16)60982-2