Cylindrical cochlear electrode array for use in humans.

Abstract

This report describes the fabrication of a flexible multichannel electrode array suitable for use in humans. The conductors, pads, and stimulating tips are made of platinum on a polyimide substrate. Photolithographic techniques are employed in the fabrication of the electrode on a planar surface in the form of a film. The film is rolled subsequently into a cylinder of diameter 0.50 mm and the cylinder is filled with medical grade silicone rubber. The stimulation pads then form rings around the cylinder. In vitro and in vivo tests are ongoing with good results so far.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

P D van der Puije, C R Pon, H Robillard. 1989. Cylindrical cochlear electrode array for use in humans.. https://doi.org/10.1177/000348948909800614

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

The perceptual effects of interphase gap duration in cochlear implant stimulation.

The most common current pulse shape used for cochlear implants is a biphasic rectangular pulse. The interphase gap (IPG) is the duration of the zero-current portion which lies between the two phases. It is known from single-nerve studies in animals that, as the IPG decreases, the biphasic pulse becomes less efficient in activating the nerve cell. Thus, it can be predicted that stimulation using smaller IPGs will necessitate the use of higher currents to maintain the loudness required by the cochlear implantee. The development of contemporary processing schemes commonly involves the maximization of the rate parameter, and to achieve this in sequential pulsatile stimulation, the IPG as well as the pulse phase duration must be minimized. This experiment investigated the effect of IPG on loudness in eight cochlear implantees who use the CI24 implant manufactured by Cochlear Ltd. An exponential increase in current level was required to maintain equal loudness when IPG is reduced from 100 to 45 and 8.4 micros. The effect of IPG was greater at lower levels, was greater for shorter pulse durations (26 micros compared to 52 micros), and was not significantly different for the rates (1 kHz or 4 kHz) tested.

Cochlear Implants↗