Abstract
Expressions describing the absorbed dose and the number of incident photons necessary for the detection of a contrasting detail in x-ray transmission CT imaging of a circular phantom are derived as functions of the linear attenuation coefficients of the materials comprising the object and the detail. A shell of a different material can be included to allow simulation of CT imaging of the skulls of small laboratory animals. The equations are used to estimate the optimum photon energy in x-ray transmission computed microtomography. The optimum energy depends on whether the number of incident photons or the absorbed dose at a point in the object is minimised. For a water object of 300 mm diameter the two optimisation criteria yield optimum photon energies differing by an order of magnitude.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
P Spanne. 1989. X-ray energy optimisation in computed microtomography.. https://doi.org/10.1088/0031-9155%2F34%2F6%2F004