Capnometers.

Abstract

The process of monitoring carbon dioxide in the respired gas is described. The physical principle employed in many capnometers is nondispersive infrared absorption: An infrared light beam is projected through a gas sample and the intensity of transmitted light is measured. Differentiation between sidestream and mainstream capnometers is made, where appropriate, in the description of this process. Important factors such as sampling flow rate, water removal, exhaust gas scavenging, measurement interference, calibration, the effect of water vapor, and measurement chamber cleaning are described as part of this process. The value of the data displayed, once derived, is based on the knowledge that carbon dioxide is an end product of the metabolism of oxygen-consuming organisms. The data can be a useful indicator of metabolic, circulatory, and ventilatory processes.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

L R Mogue, B Rantala. 1988. Capnometers.. https://doi.org/10.1007/bf01641813

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

[Experience with hydrogen (H2) breath test].

The authors applied regularly the hydrogen (H2) breath test during the medical investigations in patients with gastrointestinal symptoms. On the basis of the repeated examinations the hydrogen (H2) breath test is a sensitive, well repeatable method. After analysing of 108 examinations the authors have observed in 51.8% lactose intolerance, in 17.8% small intestinal bacterial overgrowth syndrome, in 46.4% motility disorders. The proportion of "low hydrogen producers" was 14.3%. Only 1/5 of patients with new diagnosed lactose intolerance had knowledge about the intolerance, and 10% was asymptomatic. In case of small intestinal bacterial overgrowth syndrome the repeated hydrogen (H2) test may indicate the effectivity of applied antibiotic therapy. The mean orocaecal transit time was 99 minutes and the normal range was 66-132 minutes, counted on the basis of mean +/- 2 SD. The authors suggest that the results support the important role of hydrogen (H2) breath test in the modern gastroenterological diagnostics.

Breath Tests