A practical approach to ultrasonic imaging using diffraction tomography.
Abstract
A technique for ultrasonic imaging based on the theory of diffraction tomography is presented. The method utilizes a fixed, circular configuration of transmitters and detectors. This configuration was selected because it avoids many practical limitations associated with the design of a medical imaging device. Practical considerations also motivated the inclusion of effects associated with the transmitter beam pattern rather than pursuing the more conventional approach in which plane-wave illumination is required. In addition, the problem of separately imaging both density and compressibility variations is considered.