Pulse oximetry: applications and limitations.

Abstract

The pulse oximeter estimates arterial hemoglobin saturation by measuring the light absorbance of pulsating vascular tissue at two wavelengths. The relationship between measured light absorbances and saturation was developed empirically and is built into the oximeter software. Studies in human volunteers have shown good performance of the device in healthy adults for saturations in the range of 70 to 100%. Studies in the operating room and intensive care unit have established its clinical accuracy and usefulness. The pulse oximeter has already found a number of clinical applications outside of the operating room, such as monitoring during patient transport, respiratory monitoring during narcotic administration, and evaluation of home-oxygen therapy. To use this monitor to its full potential, we must be aware of its limitations as well as its advantages. Because of the nature of the HbO2 dissociation curve, saturation measurements will not be sensitive to changes in PaO2 when the PaO2 is greater than 100 torr. This also implies that the pulse oximeter may fail to detect an inadvertent endobronchial intubation in the operating room. It may take minutes to detect an esophageal intubation in a well-preoxygenated patient. When desaturation does occur, the pulse oximeter detects it quickly, accurately, and reliably. Since the pulse oximeter uses two wavelengths of light, it cannot distinguish more than two hemoglobin species. Thus, COHb and MetHb will cause errors in SpO2 if present in large amounts. Intravenously administered dyes can also cause errors because of their absorbance properties, particularly methylene blue and indocyanine green. The pulse oximeter may be unable to detect an adequate signal during abnormal hemodynamic conditions. The pulse oximeter is one of the most important advances in noninvasive monitoring because it provides a means of continuously and quickly assessing arterial blood oxygenation. It is easy to use and interpret, requires little setup time, and poses no additional risks to the patient. Pulse oximetry may soon be a standard of practice for routine monitoring in any clinical setting in which the patient is at risk of hypoxemia.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

S J Barker, K K Tremper. 1987. Pulse oximetry: applications and limitations.. https://doi.org/10.1097/00004311-198702530-00010

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Sublingual capnometry: a non-invasive measure of microcirculatory dysfunction and tissue hypoxia.

With improvement in supportive care patients rarely die from their presenting illness but rather from its sequela, namely sequential multi-organ failure. Tissue hypoxia is believed to be the causation of multi-organ dysfunction syndrome (MODS). The expedient detection and correction of tissue hypoxia may therefore limit the development of MODS. The standard oxygenation and hemodynamic variables (blood pressure, arterial oxygenation, cardiac output) which are monitored in critically ill patients are 'upstream' markers and provide little information as to the adequacy of tissue oxygenation. Global 'downstream' markers such as mixed venous oxygen saturation and blood lactate are insensitive indicators of tissue hypoxia. Sublingual PCO(2) is a regional marker of microvascular perfusion and tissue hypoxia that holds great promise for the risk stratification and end-point of goal directed resuscitation in critically ill patients. This paper reviews the technology and application of sublingual PCO(2) monitoring.

Blood Gas Monitoring, Transcutaneous↗

Measurement of combined oximetry and cutaneous capnography during flexible bronchoscopy.

The aim of the present study was to assess the feasibility of measuring combined arterial oxygen saturation measured by pulse oximetry (Sp,O2) and cutaneous carbon dioxide tension (Pc,CO2) to monitor ventilation and quantify change in Pc,CO2 during bronchoscopy. Combined Sp,O2 and Pc,CO2 were measured at the ear lobe in 114 patients. In four patients, the ear-clip slipped and they were excluded. In total, 11 patients had artefacts with Sp,O2 recordings, thus, Sp,O2 was analysed in 99 patients. Spirometry data were available in 77 patients. Multivariate analysis of covariance and logistic regression were used for statistical analyses. Mean baseline Pc,CO2 was 4.78+/-1.06 kPa (36+/-8 mmHg) and mean rise in the Pc,CO2 during bronchoscopy was 1.26+/-0.70 kPa (9.5+/-5.3 mmHg), while mean Pc,CO2 at the end of bronchoscopy was 5.85+/-1.19 kPa (44+/-9 mmHg) . Baseline Pc,CO2 and the lowest Sp,O2 were significantly associated with peak Pc,CO2 and the change in Pc,CO2 during bronchoscopy. Risk of significant hypoxaemia (Sp,O2<or=90%) was lower for a higher baseline Sp,O2. Peak Pc,CO2 was directly associated with significant hypoxaemia. There was no significant association in the baseline Pc,CO2, peak Pc,CO2, baseline Sp,O2 or the lowest Sp,O2 comparing patients with and without chronic obstructive pulmonary disease. In conclusion, it is feasible to measure combined pulse oximetry and cutaneous carbon dioxide tension effectively to monitor ventilation during flexible bronchoscopy.

Blood Gas Monitoring, Transcutaneous↗

Transcutaneous carbon dioxide monitoring.

Technologies now exist that measure carbon dioxide levels transcutaneously. Rapid assessment of patients who have depressed ventilation or suspected sepsis can improve treatment decisions including the need for admission to the ICU and pulmonary artery catheterization.

Blood Gas Monitoring, Transcutaneous↗