PubMed · 3654228
Angular dependence of shallow dose from energetic electrons.
Abstract
A definite angular dependence of shallow dose from energetic electrons was observed when measuring the angular response of several detectors. This angular dependence of dose appeared to be fundamental, not a detector artifact. The theoretical response of a detector is discussed and compared to measurements of shallow dose with tissue and phantom response detectors. The energy-dependent angular response of the dose observed could not be explained by simple trigonometric arguments. The response, which is backscatter-dependent, must be considered in detector design and dose measurements. It is not possible for standard detectors to follow this response.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
J L Alvarez. 1987. Angular dependence of shallow dose from energetic electrons.. https://doi.org/10.1097/00004032-198710000-00007
Cite the original work for its findings. Save a collection to share your selection of sources.