PubMed · 3705276
Screening for urinary tract infection using Bac-T-Screen bacteriuria device.
Abstract
The Bac-T-Screen was used to process 795 urine specimens. Tests for urine specimens took slightly more than two minutes. The Bac-T-Screen predicted with 99 per cent accuracy if a specimen was negative for bacteriuria or pyuria. Thus, approximately one half of all carefully collected urine specimens need no further laboratory culture. In addition, the Bac-T-Screen detected bacteriuria with a sensitivity of 96 per cent at the 10(5) CFU/ml level of probability. Pyuria (1 +) was detected with a sensitivity of 98 per cent. The Bac-T-Screen can be used in an office practice as well as in the clinical laboratory.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
O P Khanna, D L Son. 1986. Screening for urinary tract infection using Bac-T-Screen bacteriuria device.. https://doi.org/10.1016/0090-4295(86)90407-3
Cite the original work for its findings. Save a collection to share your selection of sources.