PubMed · 3809472
Cardiac function: evaluation with fast-echo MR imaging.
Abstract
Magnetic resonance (MR) imaging of the heart has, to date, been limited in its ability to evaluate cardiac function. The authors have implemented a technique for functional assessment of the heart using shorter echo times than those generally used for conventional spin-echo imaging. With these short echo times, multiple images can be obtained in a multisection mode approximately within the isovolumetric phases of the cardiac cycle. This permits a pair of image stacks to be obtained, one in end systole and the other in end diastole. With the use of a modified Simpson rule, left ventricular volume and ejection fraction were calculated and compared with results obtained from contrast material-enhanced ventriculography. Preliminary results indicate that this method has promise for the evaluation of a variety of functional parameters in the heart. The short acquisition times for this functional study permit it to be combined with a tissue characterization study within the time constraints of a clinical MR imaging session.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
R R Edelman, R Thompson, H Kantor, T J Brady, M Leavitt, R Dinsmore. 1987. Cardiac function: evaluation with fast-echo MR imaging.. https://doi.org/10.1148/radiology.162.3.3809472
Cite the original work for its findings. Save a collection to share your selection of sources.