Design data for efficient axial gradient coils: application to NMR imaging.
Abstract
A new design for an axial magnetic field gradient is described. Implemented in a four-coil configuration, it requires far less power than the conventional Maxwell pair, while maintaining the same field linearity. A practical design tool with a set of curves giving coil dimensions is proposed. Two realizations dedicated to NMR imaging are described and compared with the equivalent Maxwell pair. Substantial power reductions are achieved; in these cases, dc power is reduced by a factor of 5 and switching power by a factor of 15.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
H Saint-Jalmes, J Taquin, Y Barjhoux. 1985. Design data for efficient axial gradient coils: application to NMR imaging.. https://doi.org/10.1002/mrm.1910020307