PubMed · 3835746
Lock-ring for Tenckhoff catheter/titanium adaptor junction.
Abstract
One standard continuous ambulatory peritoneal dialysis (CAPD) infusion/drainage system was found prone to recurrent mechanical failure at the junction between the silastic catheter and the titanium connector. This resulted in slackening and eventual disconnection with attendant risk of bacterial entry and a CAPD-associated peritonitis. A simple "lock-ring" device, when fitted over the catheter-titanium connector junction, afforded a secure joint.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
S Hudson, R Mackie, W K Stewart. Lock-ring for Tenckhoff catheter/titanium adaptor junction.. https://doi.org/10.3109/08860228509104842
Cite the original work for its findings. Save a collection to share your selection of sources.