A computer controlled calibrator, spectrum overlap eraser and automatic sensitivity control for an anaesthetic mass spectrometer.

Abstract

The operation, design and performance of a microcomputer controlled spectrum overlap eraser and automatic sensitivity control (ASC) for a clinical mass spectrometer are described allowing calibration for up to eight anaesthetic gases in less than 15 s per gas. The critical design parameters are identified and the mathematical algorithm used discussed. Constructional details of the system are given with particular reference to the key components identified in the analysis of design parameters. The results of static and dynamic testing of the system are reported. The calibrator showed output linearities within +/- 1.7% first standard deviation (FSD) for a two part mixture of gases relevant in anaesthesia and a typical accuracy of 2.1% FSD for any point. Dynamic testing showed that the machine did not impede the overall speed of response of the system. Transient viscosity effects were adequately eliminated and ASC action effective for variation in source pressure between 11.3 X 10(-3) Pa and 2.7 X 10(-4) Pa. Mean drift with the calibrator was found to be less than 0.38% vol/vol h-1.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

P C Beatty, J M Cooley. 1985. A computer controlled calibrator, spectrum overlap eraser and automatic sensitivity control for an anaesthetic mass spectrometer.. https://doi.org/10.1088/0143-0815%2F6%2F4%2F006

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Perception of training needs and opportunities in advanced airway skills: a survey of British and Irish trainees.

BACKGROUND AND OBJECTIVE: We surveyed delegates at the Group of Anaesthetists in Training (UK) meeting to investigate evidence of a training-gap (number of fibreoptic intubations believed to bestow competence vs. number actually performed). METHODS: Questionnaires were distributed to and collected from delegates in person. Questions covered six areas, including experience of fibreoptic intubation and cricothyrotomy, fibreoptic intubation as a specialist skill and ethical issues. RESULTS: We received 221 replies (76%). All trainees believed competence to be achievable with 10 intubations (interquartile range (IQR) 10-20); the median number performed was 2 (IQR 0-4). This was statistically significant for the groups' senior house officers, 1st and 2nd year registrars and 3rd and 4th year registrars; P < 0.0001. Many final year trainees (12/20, 60%) also failed to achieve their competency target. Few trainees had seen or performed any cricothyrotomies (medians 0, IQRs 0-1 and 0-0). Most (195/208, 94%) believed that fibreoptic intubation was a core skill and 199/212 (94%) believed that all should be competent by completion of training. Ten percent (n = 208) felt it unethical to perform an awake training intubation with full consent and 10% believed it acceptable without explanation. Most (82.7%) would fibreoptically intubate an asleep patient (requiring intubation) without consent. CONCLUSION: Trainees reported a gap between their perception of competence and achievement in awake fibreoptic intubation. Simple and complex simulations and structured training programmes may help. Anaesthetists must address the ethics of clinical training in advanced airway management.

Anesthesiology↗