An applicability study on a CT-guided stereotactic technique for functional neurosurgery.

Abstract

When a CT-guided stereotactic technique for functional neurosurgery is adopted, extremely precise targeting is needed to obtain satisfactory surgical results. In this study the authors have investigated the accuracy of the target points determined by CT-guided techniques and compared with those of conventional roentgenographically controlled stereotactic procedures. Stereotactic surgery, employing the Brown-Roberts-Wells (BRW) system, was performed contemporarily 26 times in 23 patients, that is, 9 times in 8 patients for functional neurosurgery using with the roentgenographic method, and 17 times in 15 patients with the CT-guided method only for intracranial neoplasm biopsy. As a result, there were no problems of accuracy of determining the target points by CT-guided stereotactic surgery with the BRW system. When applying this technique for functional neurosurgery, it should be pointed out that there could be a discrepancy within 2 mm from the conventional target determination.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

T Asakura, K Uetsuhara, R Kanemaru, K Hirahara. 1985. An applicability study on a CT-guided stereotactic technique for functional neurosurgery.. https://doi.org/10.1159/000101100

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

A penalized likelihood approach to magnetic resonance image reconstruction.

Currently, images acquired via magnetic resonance imaging (MRI) and functional magnetic resonance imaging (fMRI) technology are reconstructed using the discrete inverse Fourier transform. While computationally convenient, this approach is not able to filter out noise. This is a serious limitation because the amount of noise in MRI and fMRI can be substantial. In this paper, we propose an alternative approach to reconstruction, based on penalized likelihood methodology. In particular, we focus on non-linear shrinkage estimators and show that this approach achieves a great reduction in integrated mean squared error (IMSE) of the estimated image with respect to the currently used estimator. This approach is extremely fast and easy to implement computationally. In addition, it can be combined with various alternative approaches to MR image reconstruction and can be easily adapted to other, non-MRI contexts, in which the observed data and the quantities of interest are related via a linear transform.

Brain Diseases↗