PubMed · 3951414
A laser-based multiformat camera for medical imaging.
Abstract
The spatial resolution and contrast resolution required of a multiformat camera (MFC) for medical imaging are discussed. A typical cathode-ray tube (CRT) MFC and a prototype laser MFC are compared based on the following measured quantities: line spread function and associated contrast transfer function, noise characteristics, intensity transfer function (dynamic range), large-area contrast, and film irradiance. The laser MFC is found to provide significantly better performance than the CRT MFC in all of these areas.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
J N Sanders, C L Cattell, N E Bender, M M Tesic, R A Sones. A laser-based multiformat camera for medical imaging.. https://doi.org/10.1118/1.595927