Abstract

Anaesthetic equipment was studied to determine whether the accuracy was improved and failure rate decreased by routine maintenance and calibration by a biomedical technician. Each piece was evaluated, and then repaired and rechecked at intervals by the same technician. Equipment failures were divided into three types: first, equipment that was completely nonfunctional; second, equipment that was functional but inaccurate; and third, equipment that was functional and accurate but needed minor repairs. The percentage of equipment failures in each group was compared on initial evaluation and after 6 months. Of the 311 pieces of equipment, 40% needed repair at the time of the initial survey; 80% was nonfunctional, and 18% was functional but inaccurate. After six months on a maintenance schedule, only 15% of the equipment needed repair, 3% was nonfunctional, and 6% was functional but inaccurate. The difference between the total percentage of equipment failure initially and after six months was statistically significant. After a regular maintenance, calibration, and checkout schedule by a biomedical technician was instituted, there was a significant improvement in the accuracy of the equipment and a reduction in the percentage of equipment needing repair.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

H S Holley, J S Carroll. 1985. Anaesthesia equipment malfunction.. https://doi.org/10.1111/j.1365-2044.1985.tb10505.x

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Perception of training needs and opportunities in advanced airway skills: a survey of British and Irish trainees.

BACKGROUND AND OBJECTIVE: We surveyed delegates at the Group of Anaesthetists in Training (UK) meeting to investigate evidence of a training-gap (number of fibreoptic intubations believed to bestow competence vs. number actually performed). METHODS: Questionnaires were distributed to and collected from delegates in person. Questions covered six areas, including experience of fibreoptic intubation and cricothyrotomy, fibreoptic intubation as a specialist skill and ethical issues. RESULTS: We received 221 replies (76%). All trainees believed competence to be achievable with 10 intubations (interquartile range (IQR) 10-20); the median number performed was 2 (IQR 0-4). This was statistically significant for the groups' senior house officers, 1st and 2nd year registrars and 3rd and 4th year registrars; P < 0.0001. Many final year trainees (12/20, 60%) also failed to achieve their competency target. Few trainees had seen or performed any cricothyrotomies (medians 0, IQRs 0-1 and 0-0). Most (195/208, 94%) believed that fibreoptic intubation was a core skill and 199/212 (94%) believed that all should be competent by completion of training. Ten percent (n = 208) felt it unethical to perform an awake training intubation with full consent and 10% believed it acceptable without explanation. Most (82.7%) would fibreoptically intubate an asleep patient (requiring intubation) without consent. CONCLUSION: Trainees reported a gap between their perception of competence and achievement in awake fibreoptic intubation. Simple and complex simulations and structured training programmes may help. Anaesthetists must address the ethics of clinical training in advanced airway management.

Anesthesiology↗